Biological and Pharmaceutical Bulletin
Online ISSN : 1347-5215
Print ISSN : 0918-6158
ISSN-L : 0918-6158
Regular Articles
Development of Radioimmunoassay for Measurement of Serum Digoxin in Digitalized Patients Using Novel Anti-digoxin Antiserum
Yukari IkedaTsutomu ArakiHiroaki TakimotoYouichi Fujii
Author information
JOURNAL FREE ACCESS

2002 Volume 25 Issue 4 Pages 422-425

Details
Abstract
There is an antiserum elicited by digoxin 3′-hemisuccinate–bovine serum albumin (BSA) conjugate possessing high specificity for digoxin. Our study focused on development of RIA using this novel antiserum for measurement of digoxin in serum from digitalized patients. The property of the new antiserum was investigated by RIA with digoxin 3′-hemisuccinyl-[3H]leucine. The separation of bound and free fractions was performed using a dextran-coated charcoal suspension. The new antiserum bound approximately 50% of digoxin 3′-hemisuccinyl-[3H]leucine with a final dilution of 1 : 30000. The intra- and inter-assay coefficients of variation were <9% in the range of 0.52—4.17 ng/ml. The mean digoxin concentration in serum samples (n=35) from digitalized patients was estimated to be 0.68 ng/ml, which was lower than its measurement of digoxin with the commercial anti-digoxin BSA serum and monoclonal anti-digoxin. It is apparent that the RIA described here has sufficient precision. The RIA system was available for the measurement of digoxin in serum from digitalized patients.
Content from these authors
© 2002 The Pharmaceutical Society of Japan
Previous article Next article
feedback