Fructosamine Assay Using Albumin Extracted from Serum
Emi OhkawaraYukio NoharaYoshihiko KannoHiromichi SuzukiGou MatsumotoToshio KinoshitaMitsuo Watanabe
Author information
JOURNAL FREE ACCESS

2002 Volume 25 Issue 9 Pages 1121-1124

Details
Abstract

Albumin extracted from serum by a simple technique using trichloroacetic acid and ethanol has been applied to a fructosamine assay using nitroblue tetrazolium. A fructosamine assay using extracted albumin sample was carried out without interference from low molecular weight substances with reducing activities and other proteins with varying concentrations, half-lives and reducing activities. 1-Deoxy-1-p-toluidino-D-fructose (DTF) was selected as a calibrator for the assay because it is a stable Amadori reaction product with a constant glycation rate. Albumin fructosamine value was calculated using the calibration curve of DTF. The corrected albumin fructosamine value was expressed as the amount of albumin fructosamine per gram of extracted albumin taking into consideration the variation in albumin concentrations in sera from patients. The corrected albumin fructosamine values correlated more closely with the fasting blood glucose levels (r=0.735) than the serum fructosamine values corrected for albumin concentrations (r=0.514) (p<0.05).