Biological Sciences in Space
Online ISSN : 1349-967X
Print ISSN : 0914-9201
ISSN-L : 0914-9201
3D- クライノスタットとその動作特性
山下 雅道山下 明子山田 晃弘
著者情報
ジャーナル フリー

1997 年 11 巻 2 号 p. 112-118

詳細
抄録

Three dimensional clinostat has been developed for simulation of microgravity on ground. It has applied in many disciplines in gravitational biology. Outline of operational principle is described together with its mechanical design. Rotation around two independent axes makes direction of gravity vector to scan whole steric angle. Magnitude and direction of rotational angular velocity is selected randomly at a certain interval of time to avoid singularity in sweep trajectory of gravity vector. Methods for validation of the operation are presented to test randomness of motion and cancellation of gravity by clino-rotation. Concerns discussed are vibration originated in motor and pseudo-weak magnetic field generated on clinostat. Fluid flow induced by clino-rotation is pointed as another problem to be taken into account.

著者関連情報
© 1997 日本宇宙生物科学会
前の記事
feedback
Top