分析化学
Print ISSN : 0525-1931
DEVELOPMENT AND PERFORMANCE OF AN AUTOMATIC HYDRIDE/MERCURY SYSTEM
Yongming LIUBenling GONGJuhua SUNTiezheng LIN
著者情報
ジャーナル フリー

1983 年 32 巻 1 号 p. E9-E15

詳細
抄録
All operations of this system are automatically and sequentially controlled by a programmer. The calibration curve is obtained by using only one standard solution. The optimal operating parameters were found for arsenic, bismuth, lead, antimony, selen, tin and mercury. The sensitiveties and detection limits of these elements are 0.15, 0.04, 0.24, 0.05, 0.05, 0.10, 0.11 and 0.1, 0.01, 0.1, 0.02, 0.04, 0.03, 0.03 ng/ml respectively. The RSDs are lower than 4% for all elements. This system was evaluated with 35 samples of various marine products, sediment and effluent.
著者関連情報
© The Japan Society for Analytical Chemistry
前の記事 次の記事
feedback
Top