α-ニトロソ-β-ナフトール錯体として水晶重みセンサー上への付着を利用する水中の微量コバルト(II)の定量
野村 俊明湯本 さえ子
著者情報
ジャーナル フリー

1991 年 40 巻 8 号 p. 417-420

詳細
抄録

Cobalt(II) reacts with α-nitroso-β-naphthol to form an insoluble complex in water. The precipitate adsorbed on a piezoelectric quartz crystal shifted the frequency. The frequency shifts were proportional to the amounts adsorbed. Acetic acid (10 mM) and α-nitroso-β-naphthol (1.5 mM) solutions were passed through the quartz mass sensor to get constant frequency (F1). And then instead of the stream of the acetic acid solution, the cobalt(II) sample in 10 mM acetic acid solution was passed. After exactly 5 min, the stream was changed to the acetic acid solution to get a constant frequency (F2). The frequency change resulting from the adsorption of cobalt(II) complex was calculated as the difference, F1-F2. The concentration range (260) μM of cobalt(II) could be determined from the calibration curve with good reproducibility. Cobalt(II) complex adhered could be removed from the sensor with acetic acid.