Decreased Defibrillation Threshold and Minimized Myocardial Damage With Left Axilla Implantable Cardioverter Defibrillator Implantation
Mahito NoroXin ZhuYoshinari EnomotoYasuhiro OikawaHiroyuki TatsunamiRina IshiiYasutake ToyodaMasako AsamiNaohiko SaharaTakahito TakagiYuriko NarabayashiHikari HashimotoNaoshi ItoShingo KujimeTsuyoshi SakaiKeijirou NakamuraTakao SakataHaruhiko AbeKaoru Sugi
Author information
JOURNAL FREE ACCESS FULL-TEXT HTML

2016 Volume 80 Issue 4 Pages 878-886

Details
Abstract

Background: To reduce myocardial damage caused by implantable cardioverter defibrillator (ICD) shock, the left axilla was studied as an alternative pulse generator implantation site, and compared with the traditional implantation site, the left anterior chest.

Methods and Results: Computer simulation was used to study the defibrillation conduction pattern and estimate the simulated defibrillation threshold (DFT) and myocardial damage when pulse generators were placed in the left axilla and left anterior chest, respectively; pulse generators were also newly implanted in the left axilla (n=30) and anterior chest (n=40) to compare the corresponding DFT. On simulation, when ICD generators were implanted in the left axilla, compared with the left anterior chest, the whole heart may be defibrillated with a lower defibrillation energy (left axilla 6.4 J vs. left anterior chest 12.0 J) and thus the proportion of cardiac myocardial damage may be reduced (2.1 vs. 4.2%). Clinically, ventricular fibrillation was successfully terminated with a defibrillation output ≤5 J in 86.7% (26/30) of the left axillary group, and in 27.5% (11/40) of the left anterior group (P<0.001).

Conclusions: Clinically and theoretically, the left axilla was shown to be an improved ICD implantation site that may reduce DFT and lessen myocardial damage due to shock. Lower DFT also facilitates less myocardial damage, as a result of the lower shock required. (Circ J 2016; 80: 878–886)

Implantable cardioverter defibrillators (ICD) are widely used for the prevention of sudden cardiac death due to ventricular arrhythmia.1,2 A worsened prognosis has been reported, however, even after appropriate ICD shock, accompanied by a significant increase in the risk of mortality.3,4 Further studies have been done to determine whether the adverse prognosis is caused by underlying cardiac arrhythmia or the side-effect of shocks.58 Although it is controversial whether the myocardial damage caused by ICD shock worsens the prognosis,3 previous studies indicated that ICD shock causes myocardial damage, and severity is related to the defibrillation energy.913

Editorial p 816

Although the outcome of myocardial damage is controversial regarding the prognosis, it is hypothesized that the myocardial damage caused by ICD shock may have a significant adverse influence on a poorly functioning myocardium. Therefore, low-energy ICD shock, enabled by lower defibrillation threshold (DFT), may reduce the myocardial damage and thus improve the prognosis.

In addition, the electrical conduction is determined by the spatial distribution of the electrical conductivity of the tissue, and usually is not directly propagated in the human body. Because the DFT is related to the distribution of the defibrillation currents, the study of DFT should be based on the electrical conduction of defibrillation shocks.

In our previous study, pacemaker pulse generators were implanted in the left axilla and the safety, effectiveness, comfort, and cosmetic outcome of that implantation site were confirmed.14 In this study, the left axillary chest was investigated as an alternate ICD generator implantation site theoretically and experimentally, and effective defibrillation with less myocardial damage was confirmed on computer simulation.

Methods

Computer Simulation

In order to theoretically analyze the conduction of defibrillation shocks, finite element modeling of the defibrillation was conducted using SCIRun and a dataset developed by an Open Source software project of the SCI Institute’s NIH/NIGMS CIBC Center.15 The SCIRun database includes a torso model built from cardiac gated magnetic resonance imaging (MRI) of a healthy, 19-year-old volunteer.16 The spatial interval between neighboring images was 5 mm, and every image has a dimension of 256×256. The boundary nodes, surface, volume, and meshes were obtained to construct a geometrical model of the human thorax. The torso models were piecewise made heterogeneous by setting different electrical conductivities for the different tissues: bowel gas, 0.002; connective tissue, 0.220; liver, 0.150; kidney, 0.070; skeletal muscle, 0.250; fat, 0.050; bone, 0.006; lung, 0.067; blood, 0.700; and myocardium, 0.250 siemens/m.17 By modifying the torso model, the 2 models in Figures 1A,B were used to simulate the potential distribution and spatial voltage gradients during the ICD defibrillation when the ICD generator was placed in the left anterior chest and left axilla, respectively. The defibrillation energy was calculated using E=1/2×C×V2,15 where C=130 μF, and V is the defibrillation voltage when a critical mass criterion is reached. Also note that a simulation was performed with a monophasic defibrillation waveform in order to reproduce the defibrillation current conduction and avoid the computation complexity in the simulation of a biphasic defibrillation waveform. The critical mass criterion was defined as 90% of the myocardium with a voltage gradient >5 V/cm, which was the defibrillation gradient threshold.15,18 Further, we calculated the ventricular mass with a voltage gradient >30 V/cm, which was the damage gradient threshold as an index to predict any possible myocardial damage.15,19 In the Noro et al study, the Utah model was used for the computer simulation of subcutaneous ICD.20 Given that the maximum output of ICD devices is 36 J for Biotronik, 35 J for Boston Scientific, 35 J for Medtronic, 35 J for Sorin, and 40 J for St. Jude Medical, respectively, we estimated myocardial damage up to an ICD output of 40 J in the simulations.

Figure 1.

Simulations of electrical conduction pattern through the body during implantable cardioverter defibrillator (ICD) shock vs. site of ICD generator implantation: (A) left anterior chest; (B) axillary implantation. (CH) Front, left lateral, and posterior views of the simulated electric field lines for (CE) ICD generator implantation site (A); and (FH) ICD generator implantation site (B).

Clinical Experience

During September 2012 and December 2014, we recruited 30 patients, who received a new ICD or cardiac resynchronization therapy defibrillator (CRT-D), with the ICD generator implanted in the left axilla. Table 1 lists disease, gender, age, DFT, ejection fraction (EF), body mass index (BMI), and the makers of the implanted devices. The locations of the ICD leads in the right ventricle (RV) and ICD generator are shown in Figures 2A–C.

Table 1. Left Axillary Group Subject Characteristics
Patient
ID. no.
Disease Gender Age
(yeas)
DFT ≤5 EF (%) BMI (%) Drug Device Manufacturer
1 OMI F 70 Y 43 18.1 Sota BB ICD BSJ
2 OMI M 78 Y 49 20.5 Amio ICD BSJ
3 OMI M 82 Y 43 22.5 Sota ICD SJM
4 OMI M 76 N 38 27.3 Amio, BB ICD Med
5 OMI M 58 N 39 27.3 Amio ICD Bio
6 OMI M 67 Y 54 28.5 None ICD Med
7 IVF M 22 Y 71 20.0 None ICD Sorin
8 IVF M 57 Y 69 20.3 Bepri ICD SJM
9 IVF M 79 Y 69 22.9 Bepri, BB ICD Sorin
10 IVT F 76 Y 39 14.5 Bepri ICD SJM
11 IVT M 66 Y 53 20.0 Sata, Fleca ICD SJM
12 IVT F 38 Y 73 21.6 BB ICD BSJ
13 IVT M 65 Y 63 27.5 Amio ICD Sorin
14 IVT F 64 Y 68 27.6 None ICD Med
15 IVT M 65 Y 69 28.2 BB ICD Sorin
16 DCM M 75 Y 20 20.8 BB CRT-D SJM
17 DCM F 77 Y 32 23.0 BB CRT-D BSJ
18 DCM M 57 Y 28 23.5 BB CRT-D BSJ
19 DCM F 57 N 33 28.9 Amio, BB ICD Med
20 DCM M 73 Y 15 19.7 Amio BB, Mexi CRT-D BSJ
21 BrS F 20 Y 47 21.2 None ICD SJM
22 BrS M 50 Y 75 19.9 Bepri, Quni ICD SJM
23 BrS M 55 Y 64 20.5 None ICD SJM
24 Salcoidosis F 33 Y 36 14.2 Bepri CRT-D SJM
25 Salcoidosis M 47 N 28 19.6 BB ICD BSJ
26 J wave F 28 Y 77 20.5 Amio ICD Bio
27 Long QT M 67 Y 82 25.2 Mexi ICD BSJ
28 VHD M 81 Y 38 22.6 Sota, BB ICD Med
29 VHD M 67 Y 18 23.1 BB CRT-D BSJ
30 HCM F 46 Y 68 19.2 Amio ICD Sorin

Amio, amiodarone; BB, β-blocker; Bepri, bepridile; Bio, Biotronik; BMI, body mass index; BrS, Brugada syndrome; BSJ, Boston Scientific Japan; CRT-D, cardiac resynchronization therapy defibrillator; DCM, dilated cardiomyopathy; DFT, defibrillation threshold; EF, ejection fraction; Fleca, flecainide; HCM, hypertrophic cardiomyopathy; ICD, implantable cardioverter defibrillator; IVF, idiopathic ventricular fibrillation; IVT, idiopathic ventricular tachycardia; Med, Medtronic; Mexi, mexiletine; OMI, old myocardial infarction; Quni, quinidine; SJM, St. Jude Medical; Sorin, Sorin Group; Sota, sotacole; VHD, valvular heart disease; VSA, vasospastic angina.

Figure 2.

(A) Thoracic computed tomography slice of the heart. (BE) Front and lateral views on chest radiograph with implantable cardioverter defibrillator (ICD) generator implantation in the (B,C) left axilla and (D,E) left axilla. Red arrows, most direct conduction route of the defibrillation current; yellow circle, left ventricle; rectangle, ICD generator; red dashed circle, range of defibrillation current in a homogeneous medium.

To implant the ICD generator in the left axilla, we used a left axillary vein direct puncture method and a left axillary pocket. To implant a CRT-D, the lead was inserted into the subclavian vein as usual using a Penrose drain to avoid dislodgement of the left ventricle (LV) lead. After the implantation, no special postoperative instructions were given to the subjects with an ICD or CRT-D implanted in the left axilla. The details of the procedure for the left axillary ICD implantation were similar to those for the left axillary pacemaker implantation.14 Except for the subclavian vein puncture method, the method of implantation of the CRT-D generator in the left axilla was the same as that of the ICD generator.

Furthermore, during the same period, we performed new ICD implantations in the left anterior chest in 40 patients (Table 2). The locations of the ICD leads in the RV and the ICD generator are shown in Figures 2A,D,E.

Table 2. Left Anterior Chest Group Subject Characteristics
Patient
ID no.
Disease Gender Age
(yeas)
DFT ≤5 J EF (%) BMI (%) Drug Device Manufacturer
1 OMI M 76 Y 49 19.2 Amio, BB ICD BSJ
2 OMI M 77 Y 17 23.3 Amio, BB ICD BSJ
3 OMI M 81 Y 69 25.8 Amio, BB ICD BSJ
4 OMI M 70 Y 21 21.9 BB CRT-D Med
5 OMI M 73 Y 10 19.3 Bepri, BB CRT-D Med
6 OMI M 66 Y 21 20.0 Amio, BB CRT-D Med
7 OMI M 69 N 50 20.1 Sota, BB ICD Sorin
8 OMI F 55 N 60 24.2 BB CRT-D SJM
9 OMI M 69 N 35 22.4 BB CRT-D BSJ
10 OMI M 72 N 25 24.6 Amio, BB CRT-D SJM
11 OMI M 83 N 39 22.3 Amio ICD SJM
12 OMI M 82 N 23 23.6 Amio ICD Sorin
13 OMI M 75 N 40 24.6 Sota CRT-D Sorin
14 OMI M 70 N 23 20.0 Amio, BB CRT-D BSJ
15 OMI M 72 N 28 22.0 Amio ICD BSJ
16 OMI M 75 N 33 18.9 Amio, BB, Mexi ICD Med
17 OMI F 75 N 24 17.5 Amio, BB ICD Med
18 OMI M 65 N 33 22.9 None CRT-D SJM
19 HCM M 60 N 27 24.1 None CRT-D Med
20 HCM M 66 N 33 27.4 Amio, BB ICD Med
21 HCM F 51 N 55 18.6 Amio, BB CRT-D Med
22 HCM M 70 N 41 24.0 BB CRT-D SJM
23 HCM M 44 N 71 34.9 BB ICD BIO
24 IVF M 33 N 72 23.0 None CRT-D Sorin
25 IVF M 79 N 69 21.3 Amio CRT-D Med
26 IVT M 78 N 62 20.7 BB ICD SJM
27 IVT M 67 N 20 17.6 Amio ICD SJM
28 BrS M 27 Y 55 22.1 None ICD BSJ
29 BrS M 58 N 72 25.9 None ICD BSJ
30 BrS M 43 N 65 19.4 None ICD BSJ
31 DCM M 48 N 13 30.3 None CRT-D BSJ
32 DCM F 68 Y 37 22.0 BB ICD Med
33 Salcoidosis F 71 Y 24 20.9 None ICD SJM
34 Salcoidosis M 46 N 19 19.5 Sota ICD SJM
35 Salcoidosis F 79 N 28 21.3 BB CRT-D BSJ
36 Amiroidosis M 77 N 66 23.0 None ICD BSJ
37 J wave M 44 N 58 19.3 None ICD BSJ
38 Long QT M 68 Y 55 25.2 Mexi ICD BSJ
39 VHD M 77 N 40 21.8 Amio ICD Med
40 VSA M 43 Y 62 18.0 None ICD BSJ

Abbreviations as in Table 1.

During ICD testing, the initial ICD output was set to 5 J. If the ventricular fibrillation (VF) could not be terminated using an output configuration of 5 J, ICD shocks with higher energy output were used to end VF. All defibrillations were performed using a single coil configuration (RV coil to pulse generator) regardless of the type of actual lead (single or dual coil).

The devices used in the left axillary group were made by Biotronik (n=2), Boston Scientific (n=9), Medtronic (n=5), Sorin (n=5), and St. Jude Medical (n=9), respectively. Twenty-four of these devices were ICD and the other 6 were CRT-D (Table 1). In the left anterior chest group, the devices were made by Biotronik (n=1), Boston Scientific (n=15), Medtronic (n=11), Sorin (n=4), and St. Jude Medical (n=9), respectively. Twenty-four of these devices were ICD, and the other 16 were CRT-D (Table 2). For ICD made by Medtronic and St. Jude Medical, the actual output of the ICD is equivalent to the configured output because the delivered energy can be set in the ICD. In contrast, the actual output is less than the configured output in the Boston Scientific, Sorin, and Biotronik ICD because only the stored energy can be set. As a result, for Boston Scientific, Sorin, and Biotronik ICD, the output of DFT in the DFT testing was <5 J if VF were terminated by a shock of 5 J. In view of the variations in the different output configuration methods for ICD according to the manufacturer, the primary endpoint of the DFT testing was the termination of VF by a defibrillation shock output ≤5 J.

This study was approved by the ethics committee of Toho University Ohashi Medical Center, Tokyo and was in compliance with the principles of the declaration of Helsinki. The procedure for the pulse generator implantation in the left axillary region was explained in detail to the patients, and was performed only in patients who agreed and were willing to undergo this procedure. Informed consent for implantation, and for the use and publication of the data were obtained from the patients before ICD implantation.

Statistical Analysis

Chi-squared test was used for comparison of the 2 groups regarding basic disease, gender, EF, BMI, medication, DFT, and device manufacturer. T-test was used for the analysis of age. Multiple logistic regression analysis was also used to investigate influence of implantation site, basic disease, gender, age, EF, BMI, medication, ICD type (ICD or CRT-D), and manufacturer on DFT.

Results

Computer Simulation of ICD Defibrillation

The simulated defibrillation electric field lines in the front, left lateral, and posterior views based on the 2 models are shown in Figures 1C–E and Figures 1F–H, respectively. Because the lungs and bones have a lower electrical conductivity,17 regardless of the implantation site of the ICD generator, the simulated electric field lines generally circumvent the lungs and spine. When the ICD generator was placed in the left anterior chest, the electric field lines mainly traveled in the posterior and precordial regions in the upward and downward directions, respectively (Figure 1C). The upward electric field lines reached the heart via 2 pathways: 1 ran from the bottom of the heart’s posterior wall to the heart, and the other ran down to the back of the abdomen and then to the front of the abdomen, and finally ended at the heart from the bottom of the heart’s anterior wall. When the ICD generator was placed in the left axilla, the simulated electric field lines mainly traveled in 3 directions (Figure 1F, middle row). One ran upward; 1 downward; and 1 across the precordial and posterior regions to reach the heart.

Figures 3A–F uses the isopotential surfaces to illustrate the simulated defibrillation potential distribution throughout the body during ICD shocks when the ICD generator is placed in the left anterior chest and axilla, respectively. The isopotential surfaces were perpendicular to the electric field lines. The isopotential surfaces in the mediastinum of the anterior region were denser at the anterolateral wall of the RV when the ICD was implanted in the left anterior chest (Figures 3A–C). Nevertheless, the isopotential surfaces of the left axillary implantation were denser from the lateral wall of the ICD generator to the LV (Figures 3D–F), and denser in the LV (Figures 3I,J) compared with those of the left anterior implantation (Figures 3G,H). The simulated isopotential surfaces in the 2 implantation sites had similar densities in the right apical region.

Figure 3.

Simulated distribution of the defibrillation voltage within the body for (AC) left anterior chest implantation and (DF) left axillary implantation of implantable cardioverter defibrillator (ICD) generator. (G,H) Distribution of the voltage in the sagittal plane with the heart in the (G) right anterior oblique and (H) left lateral views for left anterior chest ICD generator location. (I,J) Distribution of the voltage in (I) the frontal plane and (J) an inclined plane for axillary ICD generator location. Blue, heart; red, voltage spatial difference lower than the defibrillation gradient threshold. (K,L) Distribution of myocardium with (red) voltage gradient lower than the defibrillation gradient threshold (also shown in red in GJ) in the (K) left anterior chest and (L) left axillary implantations. (M,N) Distribution of myocardium with voltage gradient (red) above the damage gradient threshold in the (M) left anterior chest and (N) left upper axillary implantations. Slices are from the apex to the base of the ventricle.

Neither the LV nor the RV masses met the critical mass criterion for the whole ventricle. In Figure 3K, the red regions have a voltage gradient below the defibrillation gradient threshold, which occurred mainly in the LV in the left anterior chest implantation. In contrast, in the left axillary implantation, voltage gradient below the defibrillation gradient threshold occurred close to the ventricular base and extended to the atria (Figure 3L).

In the anterior chest implantation, ventricular mass with voltage gradient above the defibrillation gradient threshold was larger in the RV and smaller in the LV (Figure 4A). In contrast, in the left axillary implantation, voltage gradient above the defibrillation gradient threshold had approximately the same distribution in the RV, LV, and the whole heart (Figure 4B).

Figure 4.

Relationship between implantable cardioverter defibrillator (ICD) generator implantation site, defibrillation threshold (DFT), and myocardial damage. (A,B) Relationship between ventricular mass with voltage gradient above the defibrillation gradient threshold and defibrillation voltage. (C) DFT when both ventricles, the left, or right ventricle satisfied the critical mass criterion in the left anterior chest and left axillary implantation sites, respectively. (D,E) Relationship between ICD generator defibrillation output and proportion of ventricular mass with voltage gradient above the damage gradient threshold. (F) Ventricular mass with myocardial damage vs defibrillation output equal to the DFT of the whole heart.

The DFT of the left anterior chest implantation were 9.4 J (380 V), 12.0 J (430 V), and 4.7 J (270 V) for the whole heart, LV, and RV, respectively (Figures 4A,C). The DFT of the axillary implantation were 6.2 J (310 V), 6.4 J (315 V), and 6.0 J (305 V) for the whole heart, LV, and RV, respectively, indicating a lower DFT (Figures 4B,C).

Potential for Tissue Damage

The corresponding myocardial damage areas in the simulations were localized in the RV apex region surrounding the shock coil in the left anterior chest and left axillary implantations (Figures 3M,N), where the slices were from the ventricular apex to the base. In the simulation, when the critical mass criterion was reached for the left anterior chest implantation, the proportion of ventricular mass above the damage gradient threshold was 4.2%, 0%, and 6.8% for the whole heart, LV, and RV, respectively (Figure 4D). The corresponding values were reduced to 2.1%, 0%, and 3.4% in the left axillary implantation (Figure 4E). Thus, compared with the simulated left anterior chest implantation (Figure 4F), the potential myocardial damage was better controlled in the axillary implantation.

When the ICD output was set to 40 J (maximum output) in the simulation, the myocardial damage ratios in the whole heart, LV, and RV were 16.4% vs. 19.4%, 0% vs. 3.1%, and 26.4% vs. 29.2% for the left anterior chest and left axillary implantations, respectively (Figures 4D,E). Furthermore, the LV damage region was focused in the apical area for the left axillary implantation when the shock energy of the simulation was set to 40 J.

Human Clinical Experience

The VF in 26/30 patients (86.7%) in the left axillary group was terminated by a defibrillation output configuration of 5 J (delivery or storage energy), but only 11/40 patients (27.5%) had DFT ≤5 J in the left anterior chest group (Tables 1,2). Significant difference was observed in the clinical DFT of the 2 groups, but there was no significant difference in gender, age, EF, or BMI (Table 3). On multiple logistic regression analysis, implantation site, EF, BMI, and medications had OR 25.8, 0.28, 0.27, and 0.34, respectively (Table 4). Thus, the left axillary implantation may significantly reduce DFT, but DFT may increase with reduction in EF, increasing BMI, and prescription of amiodarone.

Table 3. Subject Characteristics vs. Implantation Site
  Left axillary group Left anterior chest group P-value
Male 34 (85) 20 (67) 0.07
Age (years) 65.0±14.3 59.9±17.6 0.18
EF <40% 21 (52.5) 12 (40.0) 0.30
BMI ≤40% 10 (25.0) 7 (23.3) 0.64
DFT     <0.05*
 ≤5 J 11 (27.5) 26 (86.7)  
 >5 J 29 (72.5) 4 (13.3)  

*P<0.05. Data given as mean±SD or n (%). Abbreviations as in Table 1.

Table 4. Multivariate Indicators of Low DFT
Index β (95% CI) OR P-value
Implantation site 3.25 (1.57 to 4.94) 25.8 0.00
Disease 0.11 (−0.24 to 0.47) 1.12 0.54
Gender −0.35 (−1.96 to 1.26) 0.70 0.67
Age 0.68 (−0.05 to 1.40) 1.97 0.07
EF −1.27 (−2.81 to 0.28) 0.28 0.11
BMI −1.28 (−2.65 to 0.10) 0.27 0.07
Drug −1.07 (−2.62 to 0.47) 0.34 0.17
Device (CRT-D or ICD) −0.25 (−1.81 to 1.31) 0.78 0.75
Manufacturer −0.43 (−1.10 to 0.25) 0.65 0.21

Abbreviations as in Table 1.

After implantation, patients with ICD visit the outpatient clinic every 4 months. The patients usually had discomfort immediately after the implantation, but this gradually disappeared within 4 months. We checked chest radiograph, sensing, impedance, and so on, on follow-up visits but did not note any migration, infection or complications. Thus, no generator replacement was performed. Furthermore, because the axilla has a relatively large and deep anatomical structure and is filled with fat tissues, the ICD/CRT-D generators implanted in the left axilla were hidden in the axilla and caused no skin trouble.

Discussion

With regard to simulation results, (1) the conducted defibrillation current avoided the lungs and spine due to their high resistance; (2) the DFT in the left axillary implantation was lower than that in the left anterior chest implantation; and (3) the proportion of ventricular mass above the damage gradient threshold was lower when the ICD was implanted in the left axilla.

When the ICD generator was placed in the left anterior chest, the downward electric field lines traveled in 2 ways (Figure 1C). The main route was from the anterior and posterior mediastinum to the heart; the other came from the chest and back and went up from the abdomen to the heart via the posterior and anterior surfaces of the diaphragm.

If the electrical conduction pattern of the left anterior chest implantation (Figure 1C) turned clockwise, it would have been similar to that of the left axillary implantation (Figure 1F). Thus, ICD shocks were also conducted across the anterior and posterior regions.

Therefore, the electrical conduction in different ICD generator implantation sites may vary as to whether the electric field lines travel to the anterior or posterior regions in the bottom of the heart. It is hypothesized that the denser the electric field lines across the posterior region in the bottom of the heart, the higher the defibrillation current conducted to the LV. In contrast, the denser the electric field lines across the anterior region in the bottom of the heart, the higher the defibrillation current conducted in the RV. That is to say, the defibrillation efficiency would be raised with higher ICD defibrillation current conducted to the LV from the posterior region of the diaphragm.18

The main difference between the defibrillation electrical conduction in the left anterior and axillary implantations was the existence of a conduction pathway intersecting the anterior and posterior regions in the left axillary implantation. The conduction mainly ran from the diaphragm to the bottom, and the corresponding defibrillation currents had a higher density in the anterior region than the posterior region. This is confirmed by the fact that the simulated isopotential surfaces (Figures 3A–F) had a higher density in the anterior region of the RV but a lower density in the LV for the left anterior chest implantation. In contrast, the simulated isopotential surfaces had a higher density in the LV in the left axillary implantation. This may have led to a higher defibrillation efficiency and lower DFT.

Therefore, on the simulation, a greater ICD energy output was required if the LV met the critical mass criterion in the left anterior chest implantation. This was proven experimentally because the LV DFT in the left anterior chest implantation was 12.0 J, and 6.4 J in the axillary implantation. Furthermore, as well as the greater ICD defibrillation current conducted to the LV, the left axillary implant location may have had a more uniform defibrillation effect than the left anterior chest implant location because the LV and RV DFT were similar in the axillary implantation.

Moreover, when the ICD generator was implanted in the left anterior chest, the electric field lines pointed directly toward the RV through the front of the precordial region including the mediastinum, and traveled to the LV by passing around the back. If the conductivity in those regions was similar, the resistance was proportional to the length of the conduction route. Furthermore, the voltage between the ICD generator and lead was fixed. As a result, the defibrillation current in the LV decreased as the conduction route from the ICD generator to the LV became longer. In contrast, in the left axillary implantation, the electric field lines across the posterior regions had a shorter conduction pathway to the LV as compared with those via the precordial region. Therefore, the defibrillation currents were stronger via the posterior region to the LV. This may have been the reason for the improvement in the defibrillation effectiveness and lower DFT.

The myocardial damage ratio of the left axillary implantation, however, was larger for the maximum ICD output because the defibrillation currents are more efficiently conducted to the LV in this implantation. Nevertheless, the myocardial damage ratio of the LV was 0% when the ICD output was ≤20 J for the 2 implantations in the simulation. Therefore, for cases of low DFT, it is feasible to set the initial DFT output as 20 J, and set the maximum DFT output when the initial defibrillation fails. For cases of high DFT, in view of the increasing of DFT due to anti-arrhythmias drug treatment and the progress of cardiac disease, it is better to set the first defibrillation shock with the maximum output to achieve a successful defibrillation effect in the LV. This may not cause severe myocardial damage in the LV because the myocardial damage ratio was approximately 3% even for maximum ICD output in the simulation.

Although the association between ICD shock and late prognosis is controversial,38 myocardial damage caused by ICD shock has been studied using animal experiments and clinical troponin indices. In animal experiments, Xie et al found that the animal survival time and cardiac index decreased with increased shock energy.9 Yamaguchi et al observed that electrical shock produces myocardial injury in animals only when the myocardial perfusion is reduced.10 Hurst et al studied 49 patients and concluded that patients with a recent myocardial infarct and mean DFT ≥18 J tend to have myocardial injury with a rise in troponin.11 Davoodi et al found that DFT test shock might result in myocardial injury.12 In the Sham’a et al study, troponin elevation occurred in most of patients after inappropriate ICD shock because of lead fracture.13 Therefore, it was proven that ICD shock causes myocardial damage, and the severity of the myocardial damage is related to the defibrillation energy. It is not clear, however, whether the myocardial damage would improve over time or become chronic. Nevertheless, myocardial damage may be an important risk factor for prognosis. After all, the lesser the myocardial damage, the better the late prognosis might be, especially for patients with low cardiac function.

In human clinical experience, 86.7% of the patients with axillary implantation had DFT ≤5 J, but only 27.5% in the control group had DFT ≤5 J. Therefore, the left axillary implantation may improve the defibrillation effectiveness, with a significantly lower DFT with potentially less myocardial damage. Furthermore, lower DFT reduces energy consumption of the battery. Although this may help to reduce the size of the ICD battery, this may also reduce the defibrillation charge time due to the lower energy output, which may speed up defibrillation. As a result, in order to avoid inappropriate shock due to shorter charge time, the duration of the VT or VF sustenance in the ICD configuration may need to be increased for axillary implantation.21

The endpoint of the present DFT test was whether the VF was terminated by 5 J. This endpoint was set because DFT testing after implantation may be necessary in patients with ICD generators implanted in the left axilla, in order to confirm the DFT and whether or not the ICD energy output is sufficient.

Conclusion

Implantation of ICD generator in the left axillary fossa instead of the traditional left anterior chest was studied. On computer simulation the defibrillation current was more efficiently conducted to the LV and therefore had a better defibrillation effect using this new implantation site. Clinical experience further validated this result. Hence, left axillary fossa implantation may lead to lower DFT compared with left anterior chest implantation, and therefore prevent myocardial damage due to ICD shock. Moreover, as well as reduction in myocardial damage, left axillary fossa implantation may also be useful for patients with high DFT.

Study Limitations

We conducted a computer simulation with the hypothesis of normal myocardium and heart structure. Conduction of the current may vary due to progression of cardiac failure and pulmonary congestion. In the clinical experience, pulse generators from different manufacturers were used, which might have different capacitances, and which may have led to variation in the DFT.

We used only one torso model in the study, but the simulation results would vary depending on use of torso models of patients with congestive heart failure, chronic obstructive pulmonary disease, pectus deformity, larger size, and so on.22 In this study, we mainly studied the conduction patterns of defibrillation shock for the anterior chest and left axillary implantations, but did not consider variation of electrical conductivity due to basic diseases.

In this study, implantation of ICD generator was carried out at the preferred site of each subject. Consequently, the patients were not randomly selected, and this study was not treated as a clinical study but rather clinical experience. Clinical study is planned in the future.

We considered only the monophasic defibrillation waveform in view of the lower mathematical complexity, because the simulation of biphasic defibrillation waveform involves the cardiac ion channels. We also expect to develop a diffusion-reaction model considering the cardiac ion channels, in order to simulate the defibrillation efficiency and lower cardiac damage of the biphasic defibrillation waveform.

Acknowledgment

The authors would like to thank Kazuya Murakami (Boston Scientific Japan) for detailed product knowledge relevant to supporting this work.

Disclosures

This work was supported in part by the Japan Society for the Promotion of Science under Grants-In-Aid for Scientific Research 25330339, and the National Institute of General Medical Sciences of the National Institutes of Health under grant number P41GM103545.