Article ID: CJ-21-0742
A 52-year-old female with refractory heart failure due to dilated cardiomyopathy underwent left ventricular assist device (LVAD; HeartMate3, Abbott, Chicago, IL, USA) implantation. One year later, transthoracic echocardiography showed restored left ventricular (LV) contraction under complete opening of the aortic valve and a reduction in LV end-diastolic diameter from 64 mm (preoperative) to 48 mm under HeartMate3 support (estimated flow 2.7 L/min at 4,700 r.p.m.). The LVAD was removed by the heart team of Tokyo Women’s Medical University. Intraoperative transesophageal echocardiography showed no interventricular communication. The postoperative course was uneventful, although LV angiography performed before discharge revealed an unexpected de novo interventricular left-to-right shunt (Figure A; Supplementary Movie), with pooling of the contrast agent in the interventricular septum (IVS) shown by computed tomography (CT; Figure B,C). A retrospective review of CT images before removal of the LVAD indicated that part of the LVAD inflow cannula had partially pierced the apical portion of the IVS due to the extremely decreased LV (Figure D,E). The shunt ratio was 1.1 and no heart failure symptoms were noted clinically. The patient was discharged and followed up at an outpatient clinic.
(A) Left ventricle angiography image obtained after removal of the left ventricular assist device (LVAD) showing an interventricular shunt (arrowhead). (B,C) Coronal computed tomography (CT) images after LVAD removal showing pooled contrast agent in the interventricular septum (IVS). LV, left ventricle. (D,E) Coronal CT images during LVAD support showing the inflow cannula of the LVAD (arrow) entering the IVS (arrowhead).
We speculated that the shunt was generated because of local fragility of the IVS injured by the overhanging inflow cannula, as well as an abrupt increase in LV pressure after removal of the LVAD. These findings highlight the possibility of ventricular assist device-induced ventricular septal perforation during LVAD support in cases with an excessively small LV.
N.H. is a member of Circulation Journal’s Editorial Team.
This study has been granted an exemption from requiring ethics approval by the Institutional Review Board of Tokyo Women’s Medical University.
Supplementary Movie. Left ventricle angiography performed after removal of left ventricular assist device showed an interventricular left-to-right shunt near the apex.