Article ID: CR-21-0071
Background: A recent optical coherence tomography (OCT) registry showed that the presence of irregular protrusion (IP) after coronary stenting was a predictor of worse 1-year cardiovascular events. This study evaluated the clinical impact of OCT-detected IP after coronary stenting at ST-elevation myocardial infarction (STEMI) culprit lesions.
Methods and Results: In all, 139 consecutive STEMI patients with OCT-detected IP after stenting were analyzed retrospectively. The maximum IP angles were measured and patients with IP were divided into 2 groups (large IP, maximum IP angle ≥180°; small IP, 0°<angle<180°). The primary endpoints were cardiac death, target vessel myocardial infarction, target lesion revascularization, and stent thrombosis at 1 year after the index percutaneous coronary intervention (PCI). Of STEMI patients with IP, 51.8% had large IP. The incidence of the primary endpoints higher was higher in the large than small IP group (12.5% vs. 1.5%, respectively; P=0.018). The occurrence of plaque rupture was an independent predictor of large IP (odds ratio 4.58; 95% confidential interval 1.86–11.27; P=0.001).
Conclusions: Maximum IP angle ≥180° was an independent predictor of clinical events in STEMI patients with IP.
Globally, ST-elevation myocardial infarction (STEMI) is one of the most life-threatening cardiac events. The development of primary percutaneous coronary intervention (PCI), coronary stenting, and imaging modalities has improved patient outcomes, and patients treated with drug-eluting stents (DES) have lower rates of repeat revascularization than patients treated with bare-metal stents.1 However, DES, including newer-generation DES, are still associated with problems such as stent restenosis, stent thrombosis (ST), and adverse clinical events.1,2
Recently, an intravascular optical coherence tomography (OCT) study revealed that the presence of irregular protrusion (IP) and a small minimum stent area (MSA) at the deployed coronary stent were predictors of 1-year device-oriented clinical events, including cardiac death, target vessel-related myocardial infarction (MI), target lesion revascularization (TLR), and ST after coronary stenting.3 However, the clinical impact of IP volume has rarely been studied to predict future culprit lesion-related events after coronary stenting. This study investigated the clinical impact of OCT-detected IP angle after coronary stenting at STEMI culprit lesions.
From January 2013 to November 2016, 217 STEMI patients were treated with OCT-guided PCI at Nara Medical University Hospital. Patients with an onset to recanalization time >48 h and those with poor image quality were excluded from the study. After post-stent OCT analysis, patients without IP were also excluded from the study.
STEMI was defined as continuous chest pain lasting >30 min, arrival at the hospital within 48 h of the onset of chest pain, ST-elevation >0.1 mV in ≥2 contiguous leads or new left bundle branch block on the 12-lead electrocardiogram (ECG), and elevated cardiac markers (creatine kinase [CK]-MB or troponin T/I).
Angiographic AnalysisCoronary angiograms performed before intervention and at the end of the procedure were analyzed. Antegrade coronary flow was evaluated according to the standard Thrombolysis in Myocardial Infarction (TIMI) criteria.4
OCT Image Acquisition and AnalysisOCT imaging was performed with an OPTIS Integrated system (Abbott Laboratories, Abbott Park, IL, USA). The OCT imaging catheter (Dragonfly OPTIS; Abbott Laboratories) was advanced distal to the target lesion. Pullback was performed during continuous injection of the contrast medium through the guide catheter. OCT images were acquired automatically at a pullback rate of 36 mm/s (180 frames/s) or 40 mm/s (160 frames/s) in the OPTIS Integrated system. OCT recording was performed 5 mm proximal and 5 mm distal to the implanted coronary stents. All OCT images were analyzed by 2 experienced investigators (A.K. and T.S.) who were blinded to the patients’ information. In the case of a disagreement between the readers, a consensus reading was obtained from a third independent investigator (M.W.).
OCT analysis included baseline and post-procedural evaluations. For qualitative analysis, all cross-sectional images within the entire stent length and 5-mm peri-stent segments were analyzed. All post-stent OCT findings needed to be visible in at least 2 consecutive cross-sectional images. The morphology of the STEMI culprit lesion was evaluated immediately after successful recanalization and classified as plaque rupture, plaque erosion, or calcified nodules. Plaque rupture was defined as the presence of fibrous cap discontinuity on the lipid core or cavity formation within the plaque.5 Plaque erosion was identified by the presence of an attached thrombus overlying an intact and visualized plaque, luminal surface irregularity at the culprit lesion in the absence of a thrombus, or attenuation of the underlying plaque by a thrombus without superficial lipids or calcification immediately proximal or distal to the site of the thrombus.5,6 The calcified nodule was defined by fibrous cap disruption detected over a calcified plaque characterized by protruding calcification, superficial calcium, or the presence of substantive calcium proximal and/or distal to the lesion.5
The presence of IP, thrombus, stent edge dissection (SED), incomplete stent apposition (ISA), and MSA were evaluated after the procedure.
IP was defined as the protrusion of tissue with an irregular surface into the lumen between stent struts. Because struts are sometimes buried within the intima, we included only in-stent protrusions with a maximum height ≥100 µm for analysis.3 If IP was identified, the IP angle was measured using the center of the mass of the lumen as the angle point instead of the center of the OCT catheter according to a previous consensus document7 (Figure 1). Patients with IP were divided into 2 groups according to the maximum IP angle: large IP (maximum IP angle ≥180°) and small IP (0°<maximum IP angle≤180°). Total IP length and maximum IP height were also documented.
Measurement of maximum irregular protrusion angle. (A) Pre-stenting image of the culprit lesion after thrombectomy. Optical coherence tomography (OCT) showed the presence of plaque rupture (arrows) and lipid-rich plaque (arrowheads). (B) Post-stent OCT image. An irregular protrusion (IP) was identified in the luminal surface over the stent strut (arrowheads). (C) Measurement of the IP angle (arrow). The IP angle was measured using the center of the lumen as the angle point. In this image, the IP angle was 228°, and this was the maximum IP angle for this case.
A thrombus was defined as a mass with a diameter ≥250 µm attached to the luminal surface, stent strut, or floating within the lumen. When a thrombus could not be completely differentiated from IP, we categorized it as IP. SED was defined as disruption of the vessel luminal surface with a visible flap at 5.0-mm proximal and distal reference segments. When SED was observed, the maximum flap length was measured. In segments without a side branch, ISA was defined as separation of the inner surface of a stent strut from the inner vessel wall by a distance greater than or equal to the axial resolution of OCT plus the width of the stent strut of each stent type, including the polymer coating. For quantitative parameters of ISA, maximum strut-to-vessel wall distance was measured as maximum ISA distance.
Cross-sectional OCT images were analyzed at 1.0-mm intervals for the following parameters: minimum lumen diameter, lesion length, mean reference diameter, and MSA. According to a previous study,3 we defined small MSA as a lesion with an MSA <5.0 mm2 for DES and <5.6 mm2 for bare-metal stents.
Clinical Outcomes at the 1-Year Follow-upAll patients were scheduled for a follow-up hospital visit around 8–12 months after coronary stenting. The primary clinical endpoints were defined as device-oriented clinical endpoints (DoCE) according to pervious research 3 and 1-year culprit lesion-related events, and included cardiac death, target-vessel-related MI, TLR, and ST.3 Cardiac death was defined as death related to immediate cardiac causes or procedure-related complications. Target vessel-related MI was defined as acute MI (AMI) related with the stented vessel. TLR was defined as any revascularization procedure of the target lesion in the presence of angiographic restenosis and signs or symptoms of ischemia. ST was defined according to the Academic Research Consortium criteria.8 The parameters for DoCE were recorded as secondary endpoints. AMI was defined as cardiac biomarker elevation, with at least 1 value above the 99th percentile of the upper reference limit, and with evidence of myocardial ischemia according to any one of the following criteria: (1) symptoms of ischemia; (2) ECG changes indicative of new ischemia (ST and T wave changes or new left bundle branch block); (3) development of pathological Q waves on the ECG; or (4) imaging evidence of new loss of viable myocardium or new regional wall motion abnormality.
Quantitative analysis of IP in post-procedure OCT imaging was evaluated to identify the OCT predictors for the primary endpoints.
The incidence of individual clinical endpoints was recorded in a non-hierarchical manner, whereas the composite endpoint was analyzed in a hierarchical manner. Finally, 1-year follow-up data were obtained for all patients. The median follow-up duration was 288.8 days (interquartile range [IQR] 349.5–255.3 days).
Statistical AnalysisCategorical variables are presented as counts and percentages. The distribution of continuous variables was evaluated with the Kolmogorov-Smirnov test. Normally distributed data are reported as the mean±SD, whereas non-normally distributed data are reported as the median and IQR. Fisher’s exact test or the Chi-squared test was used for categorical variables, and Student’s t-test or the Mann-Whitney U test was used to compare continuous variables. Multivariable logistic regression analyses were performed to determine the independent predictors of a large IP. Receiver operating characteristic (ROC) curve analyses were performed to determine the best cut-off values (using the Youden Index) for maximum IP angle and the primary endpoint. Two-sided P<0.05 was considered statistically significant. All statistical analyses were performed using JMP version 14.3.0 (SAS Institute, Cary, NC, USA).
This study was approved by the Institutional Review Board (IRB) at Nara Medical University (IRB no. 1759-2) and complied with the Declaration of Helsinki Ethical Principles for Medical Research Involving Human Subjects. Informed consent was obtained from all patients.
In all, 139 patients with IP after PCI were included in this study. Patients were divided into 2 groups: those with large IP (n=72; 51.8%) and those with small IP (n=67; 48.2%). The study flowchart is shown in Figure 2. The baseline characteristics of patients in the large and small IP groups are presented in Table 1. There was no significant difference between the 2 groups.
Study flowchart. IP, irregular protrusion; OCT, optical coherence tomography; STEMI, ST-elevation myocardial infarction.
Large IP (n=72) |
Small IP (n=67) |
P value | |
---|---|---|---|
IP at index PCI | |||
Age (years) | 64.3±10.4 | 67.1±12.6 | 0.157 |
Male sex | 60 (82.7) | 55 (82.1) | 1.000 |
Hypertension | 45 (62.5) | 45 (69.2) | 0.374 |
BMI (kg/m2) | 24.8±2.9 | 24.1±3.3 | 0.256 |
Obesity | 31 (43.1) | 29 (43.3) | 1.000 |
Current smoker | 28 (43.3) | 29 (43.3) | 0.745 |
Dyslipidemia | 67 (93.1) | 59 (88.0) | 0.388 |
Diabetes | 21 (29.2) | 25 (37.3) | 0.368 |
Insulin user | 4 (5.6) | 4 (6.0) | 1.000 |
Family history of CAD | 15 (20.8) | 14 (20.9) | 1.000 |
Prior MI | 2 (2.8) | 2 (3.0) | 1.000 |
Prior PCI | 4 (5.6) | 3 (4.5) | 1.000 |
CKD | 16 (22.2) | 18 (26.8) | 0.558 |
Dialysis | 0 (0) | 3 (4.5) | 0.109 |
Atrial fibrillation | 3 (4.2) | 2 (3.0) | 1.000 |
Laboratory data on admission | |||
Total cholesterol (mg/dL) | 191.4±42.7 | 182.0±42.4 | 0.198 |
LDL-C (mg/dL) | 125.0±44.9 | 112.0±35.9 | 0.066 |
HDL-C (mg/dL) | 45.7±11.3 | 46.9±12.4 | 0.564 |
HbA1c (%) | 6.4±1.1 | 6.3±1.0 | 0.583 |
Peak CK (IU/L) | 3,607±3,193 | 3,109±2,521 | 0.313 |
Peak CK-MB (IU/L) | 325±274 | 303±285 | 0.639 |
SCr (mg/dL) | 0.87±0.30 | 1.18±1.40 | 0.075 |
eGFR (mL/min/1.73 m2) | 76.8±25.2 | 70.1±24.0 | 0.116 |
LVEF (%) | 59.2±9.4 | 57.4±9.9 | 0.269 |
Medication on admission | |||
Aspirin | 10 (13.9) | 9 (13.4) | 1.000 |
Clopidogrel | 1 (1.4) | 0 (0) | 1.000 |
Prasugrel | 0 (0) | 0 (0) | NA |
DAPT | 0 (0) | 0 (0) | NA |
Oral anticoagulation | 0 (0) | 0 (0) | NA |
RAS blocker | 19 (26.4) | 25 (37.3) | 0.203 |
β-blocker | 4 (5.6) | 3 (4.5) | 1.000 |
Calcium channel antagonist | 19 (26.4) | 23 (34.3) | 0.358 |
Statin | 16 (22.2) | 11 (16.4) | 0.402 |
Patients with IP | |||
Angiographic characteristics | |||
Baseline TIMI flow Grade 0–2 | 60 (83.3) | 58 (86.6) | 0.642 |
Final TIMI flow Grade 3 | 57 (79.2) | 51 (76.1) | 0.689 |
ACC/AHA lesion Class B2/C | 60 (83.3) | 60 (89.6) | 0.330 |
Culprit vessel | 0.173 | ||
Left anterior descending artery | 31 (43.1) | 39 (58.2) | |
Left circumflex artery | 4 (5.6) | 4 (6.0) | |
Right coronary artery | 37 (52.4) | 24 (35.8) | |
Procedural characteristics | |||
No. stents | 0.996 | ||
1 | 61 (84.7) | 55 (82.1) | |
2 | 10 (13.9) | 11 (16.4) | |
3 | 1 (1.4) | 1 (1.5) | |
Multiple stents | 11 (15.3) | 10 (14.9) | 0.996 |
Stent type | 0.546 | ||
Bare-metal stent | 3 (4.2) | 3 (4.5) | |
Sirolimus-eluting stent | 6 (7.1) | 4 (5.0) | |
Everolimus-eluting stent | 56 (66.7) | 55 (68.8) | |
Zotarolimus-eluting stent | 14 (16.7) | 10 (12.5) | |
Biolimus-eluting stent | 5 (6.0) | 8 (10.0) | |
Stent diameter (mm) | 3.18±0.45 | 3.03±0.38 | 0.041* |
Total stent length (mm) | 26.0±10.8 | 26.1±13.2 | 0.474 |
Thrombectomy | 67 (93.1) | 58 (86.6) | 0.263 |
Distal protection | 15 (20.8) | 4 (6.0) | 0.013* |
Post dilatation | 61 (84.7) | 55 (82.0) | 0.746 |
Onset to recanalization time (min) | 240.5±141.3 | 304.3±180.5 | 0.021* |
*Indicate significant difference. Categorical variables are shown as the number (percentage) and continuous variables are shown as the mean±SD. ACC, American College of Cardiology; AHA, American Heart Association; BMI, body mass index; CAD, coronary artery disease; CK, creatine kinase; CKD, chronic kidney disease; DAPT, dual antiplatelet therapy; eGFR, estimated glomerular filtration rate; HDL-C, high-density lipoprotein cholesterol; LDL-C, low-density lipoprotein cholesterol; LVEF, left ventricular ejection fraction; MI, myocardial infarction; RAS, renin-angiotensin system; SCr, serum creatinine.
Stent diameter was significantly larger in the large IP than small IP group (3.18±0.45 mm vs. 3.03±0.38 mm; P=0.041). Thrombectomy was performed in most patients. The incidence of distal protection was significantly higher in the large IP than small IP group (20.8% vs. 6.0%; P=0.013). The onset to recanalization time was significantly shorter in the large IP group than small IP group (240.5±141.3 min vs. 304.3±180.5 min; P=0.021; Table 1).
OCT FindingsThe large IP group had a significantly higher incidence of plaque rupture (87.5% vs. 61.2%; P<0.001), but a significantly lower incidence of plaque erosion (8.3% vs. 28.4%; P=0.004) than the small IP group.
The large IP group had a significantly greater maximum IP angle (260.8±68.1° vs. 124.5±34.2°; P<0.001), longer total IP length (13.0±6.7 mm vs. 9.1±5.9 mm; P<0.001), and greater maximum IP height (0.32±0.11 mm vs. 0.23±0.08 mm; P<0.001) than the small IP group. However, the incidence of SED was lower in the large IP than small IP group (20.8% vs. 38.8%; P=0.026; Table 2).
Large IP (n=72) |
Small IP (n=67) |
P value | |
---|---|---|---|
Baseline OCT findings | |||
Lesion measurement | |||
Mean reference diameter (mm) | 3.04±0.57 | 2.96±0.66 | 0.456 |
Minimal lumen diameter (mm) | 1.09±0.40 | 1.09±0.31 | 0.989 |
Lesion length (mm) | 21.6±10.9 | 20.9±10.4 | 0.674 |
Culprit morphology | |||
Plaque rupture | 63 (87.5) | 41 (61.2) | <0.001* |
Plaque erosion | 6 (8.3) | 19 (28.4) | 0.004* |
Calcified nodule | 2 (2.8) | 7 (10.5) | 0.088 |
Post-stent OCT findings | |||
IP | |||
Maximum IP angle (°) | 260.8±68.1 | 124.5±34.2 | <0.001* |
Total IP length (mm) | 13.0±6.7 | 9.1±5.9 | <0.001* |
Maximum IP height (mm) | 0.32±0.11 | 0.23±0.08 | <0.001* |
Thrombus | 65 (90.3) | 59 (88.1) | 0.787 |
SED | 15 (20.8) | 26 (38.8) | 0.026* |
SED length (mm) | 1.00±0.55 | 0.91±0.60 | 0.330 |
ISA | 52 (72.2) | 50 (74.6) | 0.848 |
Maximum ISA distance (mm) | 0.33±0.13 | 0.36±0.16 | 0.263 |
MSA (mm2) | 6.68±2.61 | 5.96±2.47 | 0.096 |
Small MSA | 19 (26.4) | 28 (41.8) | 0.073 |
*Indicate significant difference. Categorical variables are shown as the number (percentage) and continuous variables are shown as the mean±SD. IP, irregular protrusion; ISA, incomplete stent apposition; MSA, minimal stent area; SED, stent edge dissection; OCT, optical coherence tomography.
Intra- and interobserver reliability was assessed using the kappa statistic in OCT analysis. The intra- and interobserver variabilities of the IP angle using the intraclass correlation coefficient were 0.960 and 0.886, respectively.
Clinical Events at the 1-Year Follow-upPrimary endpoints occurred in 12.5% of patients in the large IP group, compared with 1.5% of patients in the small IP group (P=0.016; Figure 3). Patients in the large IP group tended to have a higher incidence of cardiac death, target vessel-related MI, TLR, and ST compared with patients in in the small IP group; however, the differences did not reach statistical significance (Table 3). Representative OCT images of 2 stent thromboses are shown in Supplementary Figure 1. Target vessel-related MI was observed in only 1 case in the large IP group during the follow-up period.
Kaplan-Meier survival curve for the primary endpoint in patients with large and small irregular protrusions (IPs). ST, stent thrombosis; TLR, target lesion revascularization.
Large IP (n=72) |
Small IP (n=67) |
P value | |
---|---|---|---|
Primary endpoints | 9 (12.5) | 1 (1.5) | 0.018* |
Secondary endpoints | |||
Cardiac death | 4 (5.6) | 1 (1.5) | 0.367 |
Target vessel-related MI | 1 (1.4) | 0 (0) | 1.000 |
TLR | 4 (5.6) | 1 (1.5) | 0.367 |
Stent thrombosis | 2 (2.8) | 0 (0) | 0.497 |
*Indicate significant difference. Categorical variables are shown as the number (percentage). IP, irregular protrusion; MI, myocardial infarction; TLR, target lesion revascularization.
Differences in primary endpoints among patients with large IP, small IP, and without IP are presented in Supplementary Table 1.
Medications at the 1-year follow-up are listed in Supplementary Table 2. There was no significant difference between the 2 groups (Large IP vs. Small IP).
Predictors of Primary EndpointsUnivariate analysis demonstrated that large IP (P=0.009), peak CK (P=0.009), statin use at follow-up (P=0.014), left ventricular ejection fraction (P=0.042), and stent diameter (P=0.035) were correlated with the primary endpoints (Table 4).
OR | 95% CI | P value | |
---|---|---|---|
Male sex | 1.16 | 0.24–5.62 | 0.849 |
Age (per 5 years) | 0.83 | 0.66–1.04 | 0.118 |
BMI (per 5 kg/m2) | 1.09 | 0.90–1.32 | 0.245 |
Dyslipidemia | 1.90 | 0.37–9.68 | 0.464 |
CKD | 1.87 | 0.39–8.90 | 0.402 |
LVEF (per 5%) | 0.75 | 0.57–0.99 | 0.042* |
Hemoglobin | 1.35 | 0.91–2.00 | 0.105 |
LDL-C (per 10 mg/dL) | 1.14 | 0.99–1.30 | 0.062 |
Peak CK (per 100 IU) | 1.02 | 1.01–1.04 | 0.009* |
Distal protection | 3.29 | 0.600–10.13 | 0.238 |
Stent diameter (per 0.5 mm) | 2.03 | 1.04–3.99 | 0.035* |
Plaque rupture | 1.13 | 0.29–4.38 | 0.853 |
Large IP | 5.86 | 1.25–27.52 | 0.009* |
IP length (per 5 mm) | 1.22 | 0.83–1.79 | 0.586 |
SED | 0.69 | 0.18–2.67 | 0.704 |
Incomplete stent apposition | 2.11 | 0.45–10.0 | 0.311 |
Small MSA | 0.55 | 0.15–2.14 | 0.376 |
Lesion length (per 5 mm) | 0.98 | 0.74–1.29 | 0.872 |
Final TIMI flow Grade 0–2 | 2.40 | 0.73–7.96 | 0.165 |
Statin at follow-up | 0.15 | 0.04–0.61 | 0.014* |
*Indicate significant difference. CI, confidence interval; OR, odds ratio. Other abbreviations as in Tables 1,2.
The univariate model showed that large IP was positively correlated with stent diameter (P=0.040) and plaque rupture (P<0.001). Higher low-density lipoprotein cholesterol (LDL-C) levels tended to be correlated with large IP (P=0.062). However, the multivariate model revealed plaque rupture was an independent predictor for large IP (odds ratio 4.58, 95% confidence interval 1.86–11.27, P=0.001; Table 5).
OR | 95% CI | P value | |
---|---|---|---|
Univariate analysis | |||
Age per (5 years) | 0.89 | 0.77–1.04 | 0.152 |
Male sex | 1.09 | 0.45–2.63 | 0.846 |
BMI per (5 kg/m2) | 1.37 | 0.78–2.35 | 0.248 |
Dyslipidemia | 1.82 | 0.56–5.86 | 0.311 |
LDL-C (per 10 mg/dL) | 1.08 | 0.99–1.18 | 0.062 |
Statin at admission | 1.45 | 0.63–3.41 | 0.386 |
Stent diameter (per 0.5 mm) | 1.52 | 1.01–2.28 | 0.040* |
Total stent length (per 5 mm) | 0.92 | 0.87–1.14 | 0.947 |
Plaque rupture | 4.43 | 1.89–10.43 | <0.001* |
Multivariate analysis | |||
Plaque rupture | 4.58 | 1.86–11.27 | 0.001* |
Stent diameter (per 0.5 mm) | 1.52 | 0.99–2.36 | 0.052 |
LDL-C (per 10 mg/dL) | 1.07 | 0.98–1.17 | 0.144 |
*Indicate significant difference. Abbreviations as in Tables 1,4.
ROC curve analysis showed that the best cut-off value to predict primary endpoints was a maximum IP angle ≥180.0° (area under the curve=0.74; Supplementary Figure 2).
To the best of our knowledge, this is the first report of the relationship between IP volume and clinical events in patients with STEMI. The main findings of this study were that: (1) IP was frequently seen at the STEMI culprit lesion; (2) a maximum IP angle ≥180° was found in almost half the STEMI culprit lesions; (3) the incidence of primary endpoints at the 1-year follow-up was significantly higher in the large IP than small IP group; (4) large IP and statin use at follow-up were strong predictors for clinical events at 1 year after the procedure; and (5) large IP was associated with high serum LDL-C levels, larger stent diameter, and an increased incidence of plaque rupture at the STEMI culprit lesion.
Incidence of IPIP was found in 80% of STEMI culprit lesions after coronary stenting. This incidence is consistent with that in a previous study, which found that plaque rupture occurred in 67% of STEMI culprit lesions6 In addition, the incidence of large IP was 51.8% in the present study. Plaque rupture site have high incidence of large lipid core,6 and this is one of the predictors of IP;9 therefore, it is not surprising that approximately half the STEMI culprit lesions with IP had a large IP after coronary stenting.
It is important to investigate ways in which IP volume can be reduced. There is no promising device to reduce IP volume. However, excimer laser coronary atherectomy or directional coronary atherectomy may reduce IP volume by reducing the lipid core volume. It is also unknown whether reducing IP volume can prevent future cardiac events. Because the IP is generated by a following penetration of lipid plaques with a large lipid core by stent struts, aggressive medical intervention may be one way to reduce lipid volume, large IP, and the rate of clinical events.
Large IP and Clinical Events at 1 YearTo the best of our knowledge, no previous OCT study has evaluated the correlation between the size of the IP and clinical events. One qualitative OCT study demonstrated a correlation between IP and DoCE and proposed that IP may be an indicator of deep vessel injury.3 That study reported that the incidence of DoCE and TLR at 1 year was 6.5% and 5.8%, respectively.3 In contrast, in the present study, patients with large IP had a much higher incidence of DoCE (12.5%) at the 1-year follow-up. The presence of large IP may indicate the presence of deep vessel injury after coronary stenting. Cardiac death and TLR were the most common clinical events seen at the 1-year follow-up. Because the presence of IP was correlated with the presence of lipid-rich plaque before stenting,8 a large IP may be caused by a large lipid core. In addition, our study showed that plaque rupture, precursors of which are lipid-rich plaques or thin-capped fibroatheromas (TCFA), was correlated with large IP. In previous OCT studies that included only patients with acute coronary syndrome (ACS), patients with OCT-detected plaque rupture generated from lipid-rich plaque and TCFA in culprit lesions had worse prognoses or clinical events that those with erosion or an intact fibrous cap after PCI.10,11 This may be due to lipid core penetration by stent struts, which induces accelerated arterial inflammatory changes and neointimal growth.12,13 Lesions with large IP potentially result in thicker neointimal growth than those with small IP. The correlation between pre-stenting lipid-rich plaques in culprit lesions and TLR remains unclear. A study using near-infrared spectroscopy showed no correlation between baseline lipid size and clinical events after PCI.14 The reason for this may be that the study failed to show lipid injury after PCI.
The PROSPECT (Providing Regional Observations to Study Predictors of Events in the Coronary Tree) study, which used virtual histology intravascular ultrasound (IVUS), evaluated the incidence of clinical events in ACS patients who underwent coronary intervention.15 Culprit lesion-related clinical events were observed in only 12.9% of ACS patients during the 3-year follow-up period. Compared with the PROSPECT study, STEMI patients with a large IP showed a much higher incidence of culprit lesion-related events (12.5%) at the 1-year follow-up in the present study, although 30% of patients included in the PROSPECT study had STEMI. Therefore, the presence of a large IP may be a good surrogate marker to predict future culprit lesion-related events.
There were 2 cases (2.8%) of ST in patients with large IP in the present study; no cases of ST were seen in patients with small IP. Although there was no significant difference in the incidence of ST between the 2 groups, the incidence of ST in the present study was higher than that reported in a previous study.16 In another study, lesions with ST had a higher incidence of IP angles >180° in post-stent OCT findings at the index PCI compared with the control group, which included patients without ACS for at least 5 years from the index PCI.17 The presence of large IP may also be a predictive marker of ST during the 1-year follow-up.
Predictors of Large IPThe first OCT-detected IP study conducted showed that IP was more frequently observed in AMI patients.3 After this initial study, several studies revealed other predictors of IP. For example, Bryniarski et al reported obesity, AMI presentation, high serum LDL-C concentrations, percentage atheroma volume, and stent length as predictors of IP.9 Sanuki et al showed that TCFA, lipid-rich plaque, positive remodeling, and thrombus were predictors of IP.18 Considering that most patients in these studies had stable coronary artery disease, the strongest predictors of IP were ACS-related factors, such as AMI presentation, the presence of TCFA, and the presence of a thrombus. Because all patients in the present study were STEMI, we could dem- onstrate the worst prognostic type of IP and predictors of large IP. The predictors of large IP found in this study were plaque rupture, mean stent diameter, and serum LDL-C concentration. Plaque rupture was usually caused by TCFA, including lipid-rich plaque and a large plaque burden.19 Other etiologies of STEMI, such as plaque erosion and calcified nodules, rarely presented with TCFA or large lipid plaques in the culprit lesion; therefore, it was considered that lipid penetration by stent struts generated large IP. Mean stent diameter was significantly larger in the large IP group. This may be due to vessel diameter, because the lesions with large IP had a larger vessel diameter and larger plaque atheroma volume, allowing the larger stent to easily penetrate the lipid core. The OPINION (OPtical frequency domain imaging versus INtravascular ultrasound in percutaneous coronary interventiON) study compared post-stent OCT findings between IVUS- and OCT-guided PCI and showed a larger stent area and higher incidence of IP in IVUS-guided PCI, although the reference vessel areas were similar between the 2 groups.20,21 Therefore, optimal stent sizing with imaging modalities is important to avoid the occurrence of large IP.
This study also demonstrated that a high serum LDL-C concentration was a discriminator for large IP. High serum LDL-C concentrations are associated with the presence of TCFA and lipid-rich plaque,22 whereas a decrease in serum LDL-C concentration was shown to increase the thickness of the fibrous cap on the lipid core and stabilize intimal inflammation.23–25 Hence, large IP may be prevented by early introduction of lipid-lowering agents. Nishiguchi et al reported that the early introduction of statins in ACS patients resulted in a rapid increase in fibrous cap thickness in non-culprit lesions compared with patients in whom statins were introduced later.26
Study LimitationsThis study has several limitations. First, this study was retrospective in nature and was conducted at a single center that enrolled a limited number of patients. Second, the treatment strategy and the OCT procedure were performed at the discretion of each interventionist. Therefore, potential selection bias could not be avoided. Third, our classification of IP in this study is novel, and we have not investigated the correlation between OCT-detected IP and histopathological findings. Recently, combined OCT and angioscopic analysis for tissue protrusion revealed that yellow plaque was observed at a significantly higher rate in OCT-detected IP.27 Fourth, it is often challenging to discriminate thrombus from protruding tissue, including IP, that is generated from a necrotic core using OCT. Previous research suggested a close association between thrombus and IP.9 In addition, in the first report of IP, when a thrombus could not be completely differentiated from IP the authors categorized it as IP,3 and we used the same criteria in the present study. Therefore, IP and thrombus were not completely discriminated. Fifth, we reported an association between large IP and worse clinical prognosis; however, the methods to reduce IP volume remain unknown. While stenting with a small-diameter stent may reduce the incidence of large IP, a minimum stent area <5.0 mm2 in DES is also a predictor of DoCE.
In conclusion, IP was frequently observed in post-stent OCT imaging during primary PCI to manage STEMI culprit lesions. The maximum IP angle was an independent predictor of culprit lesion-related clinical events. The clinical impact of maximum IP angle in STEMI patients requires further investigation.
The authors thank the staff in the catheterization laboratory at Nara Medical University for their excellent assistance. The authors also thank Editage for the English language editing of this manuscript.
This study did not receive any funding.
The authors have no conflicts of interest to declare for this study.
This study was approved by the Ethics Committee of Nara Medical University (Approval no. 1759-2).
The deidentified participant data will not be shared.