IEICE Communications Express
Online ISSN : 2187-0136
ISSN-L : 2187-0136
Regular Section
Comparison of machine learning and non-machine learning methods for the sleep apnea detection using millimeter-wave radar
Takato KodaTakuya SakamotoShuqiong WuShigeaki OkumuraHirofumi TakiSatoshi HamadaSusumu SatoKazuo Chin
著者情報
ジャーナル フリー

2022 年 11 巻 6 号 p. 355-360

詳細
抄録

This study investigated the performance of radar-based detection methods for sleep apnea, by comparing a machine learning approach and a non-machine learning approach. We applied a 79-GHz millimeter-wave multiple-input and multiple-output array radar with 12 virtual array elements and performed radar and polysomnography measurements simultaneously to monitor a sleep apnea patient during overnight inspection in a hospital setting. Both radar-based methods successfully estimated the number of apnea events per hour of sleep, with root-mean-square error values of 4.1 and 4.0, indicating that the two methods had comparable accuracy in the radar-based noncontact monitoring of a sleep apnea patient.

著者関連情報
© 2022 The Institute of Electronics, Information and Communication Engineers