Dental Materials Journal
Online ISSN : 1881-1361
Print ISSN : 0287-4547
ISSN-L : 0287-4547
Original Papers
Application of haptic device to implant dentistry — accuracy verification of drilling into a pig bone
Takafumi OHTANINaoki KUSUMOTOKazumichi WAKABAYASHIShinichi YAMADATakashi NAKAMURAYouichi KUMAZAWAHirofumi YATANITaiji SOHMURA
著者情報
ジャーナル フリー

2009 年 28 巻 1 号 p. 75-81

詳細
抄録
To enable accurate implant placement and precise drilling following preoperative simulation, we developed the BoneNavi system. To realize more precise drilling when the holes are upsized, two methods of surgical guiding were attempted in the present study. One involved using multiple surgical guides with titanium tubes of different diameters, and the other involved using a single surgical guide but employing titanium drill guide tubes with different diameters. Drilling accuracy of the two newly developed methods was examined and compared with the results of drilling into a pig bone using only the initial surgical guide. Deviations of the position and angle with the two novel methods were similar: 0.17 mm and 1° respectively. As for the control group whereby drilling was done using only the initial surgical guide, the deviations were 0.25 mm and 3.50° — which were significantly larger than those achieved with the two novel methods. In light of the results obtained, our newly developed BoneNavi system is especially applicable for severe clinical cases that require precise implant placement.
著者関連情報
© 2009 The Japanese Society for Dental Materials and Devices
前の記事 次の記事
feedback
Top