Dental Materials Journal
Online ISSN : 1881-1361
Print ISSN : 0287-4547
ISSN-L : 0287-4547
Original Paper
Dimensional changes in stone casts resulting from long-term immersion of addition-type silicone rubber impressions in disinfectant solutions
Hisako HIRAGUCHIMasahiro KAKETANIHideharu HIROSEHisaji KIKUCHITakayuki YONEYAMA
Author information
JOURNAL FREE ACCESS

2013 Volume 32 Issue 3 Pages 361-366

Details
Abstract

There is a concern that long-term immersion of impressions in disinfectant solutions may cause changes in the dimensions of the resulting stone casts. This study investigated the dimensional changes in stone casts resulting from immersion of five brands of addition-type silicone rubber impressions in disinfectant solutions for 30 min and 24 h. Impressions of a master cast designed to simulate an abutment tooth were immersed in 2% glutaraldehyde and 0.55% ortho-phthalaldehyde. The diameter of the stone cast was measured using a laser scan micrometer. For four brands of impression materials, 30-min immersion in disinfectant solutions produced no dimensional changes in the stone casts. For four brands of impression materials, 24-h immersion caused a significant decrease in the stone cast dimensions.

Content from these authors
© 2013 The Japanese Society for Dental Materials and Devices
Previous article Next article