Mechanical effect of different patterns for preparation of orthodontic appliances: An experimental study
Shiyu TIANJuan DAIWaheed AHMADKanhui LIANGQiyun TANHao DENGFeifei JIANGJing XIONGJun CAOZeyang XIA
著者情報
ジャーナル オープンアクセス

2024 年 43 巻 5 号 p. 637-642

詳細
抄録

Archwire bending is the key to orthodontic treatment, and multi-time bendings are inevitable during manual and robotic automated bending. The purpose of this paper is to quantitatively evaluate the mechanical effects of the different preparation modes and to compare the mechanical properties of the orthodontic loops in one and multiple bends. Three types of typical stainless steel orthodontic loops (vertical loop, T-loop, and L-loop) were used to quantify the mechanical effect of patterns for preparation by experimental comparison between loops with different bending times by using an orthodontic force tester (OFT). The results were statistically analyzed by t-test. The fracture test of the stainless steel archwire was also carried out, and the bending times at fracture were recorded. Results of the tests indicate that one-time and multi-time bending have a significant mechanical effect on orthodontic appliances. Multi-time bending causes significant mechanical decreases and can damage the appliances.

著者関連情報
© 2024 The Japanese Society for Dental Materials and Devices

This is an open access article under the CC BY license
https://creativecommons.org/licenses/by/4.0/