マクロオートラジオゲラフ定量のための諸問題
泉二 奈緒美羽山 恵美子羽鳥 泰彦宍戸 亮重松 昭世
著者情報
ジャーナル フリー

1988 年 3 巻 2 号 p. 181-189

詳細
抄録
The relationship between photographic density with radioactivity of particle emitters and exposure period was studied for quantification of macroautoradiography.
Photographic density was recorded over 3 different types of X-ray high sensitive films with time by contacting with 7 different radioactivity sheets of autoradiographic standard kits, 14C-polymethacrylate.
A good correlation was obtained between radioactivity and photographic density for during a 2 or 3 day exposure period. There was also a good relationship between exposure period and photographic density in case of low radioactivity of 0.025-0.75 μCi/g in the sheets. This relationship was disturbed by unequivalent combination of the radioactivity and exposure period.
Discussion was presented to clarify the analysis of such a complex relations as mentioned above.