Fisheries science
Print ISSN : 0919-9268
Enzyme sensor system for determination of total cholesterol in fish plasma
HIDEAKI ENDOMASASHI MAITAMIO TAKIKAWAHUIFENG RENTETSUHITO HAYASHINAOTO URANOKOHJI MITSUBAYASHI
著者情報
ジャーナル フリー

2003 年 69 巻 6 号 p. 1194-1199

詳細
抄録

An enzyme biosensor system was developed for rapid and continuous determination of total cholesterol in fish plasma. The system was constructed with an immobilized enzyme membrane, Clark-type oxygen electrode, flow system and personal computer. The enzyme membrane was prepared from cholesterol esterase, cholesterol oxidase, photosensitive polymer and an ultra-thin dialysis membrane. The optimum assay conditions of the sensor system were established as follows: pH 6.8, temperature 25°C and flow rate 0.2 mL/min. The calibration curves were linear in the range of 15-80 mg/dL for yellowtail, 17-42 mg/dL for amberjack and 20-70 mg/dL for rainbow trout. One assay was completed within 5 min and it was possible to determine the sample continuously. The sensor system was applied to the determination of total cholesterol in yellowtail plasma. A good correlation was observed between the values determined by the sensor method and the conventional method (correlation coefficient was 0.90).

著者関連情報

この記事は最新の被引用情報を取得できません。

© The Japanese Society of Fisheries Science
前の記事 次の記事
feedback
Top