FORMA
Online ISSN : 2189-1311
Print ISSN : 0911-6036
Original Paper
Indirect Skin Contact Electrocardiogram Monitoring System Using Flexible Capacitive Electrodes
Issey TakahashiKiyoko YokoyamaMarco Knaflitz
著者情報
ジャーナル フリー

2017 年 32 巻 2 号 p. S19-S23

詳細
抄録

It is gaining great importance to develop and integrate physiological sensors into daily healthcare. Especially, electrocardiogram (ECG) is useful for the disease diagnosis and daily health monitoring, also for the driver's cognitive state monitoring. Capacitive ECG monitoring system, which enables to collect ECG with indirect skin contact, is implemented in this study. Signal attenuation and noise contamination is considered by installing flexible electrode, neutralization circuit, AC coupling circuit, Driven right leg (DRL), and DSP board in the configuration. The system is evaluated with 5 subjects with a car in situations of engine off and on. R-wave Detection Sensitivity (Se) and Positive Predictivity (P+) of 99.8% and 100% were obtained during the engine on.

著者関連情報
© 2017 Society for Science on Form, Japan