GEOCHEMICAL JOURNAL
Online ISSN : 1880-5973
Print ISSN : 0016-7002
ISSN-L : 0016-7002
18O/16O and D/H ratio determinations for small amounts of water
Takeshi Ohba
著者情報
ジャーナル フリー

1987 年 21 巻 4 号 p. 183-186

詳細
抄録
This paper describes an improved procedure for D/H and 18O/16O ratio analyses presented originally by Kishima and Sakai (1980). A sealed glass tube for the isotope exchange equilibration between water and CO2 gas was employed instead of a conventional container with a greased stopcock. The use of the sealed tube enabled us to handle 5 ∼ 6μmol of water for accurate and precise δ18O measurement, with precision of 0.42‰ (1 σn-1). However, for precise δD measurement more than 40μmol of water was required.