波紋
Online ISSN : 1884-636X
Print ISSN : 1349-046X
ISSN-L : 1349-046X
サイエンス記事
μPIC(Micro-Pixel Chamber)を用いた中性子イメージング検出器の開発
パーカー ジョセフ ドン原田 正英服部 香里岩城 智株木 重人岸本 祐二窪 秀利黒澤 俊介松岡 佳大身内 賢太朗水本 哲矢西村 広展奥 隆之澤野 達哉篠原 武尚鈴木 淳市高田 淳史谷森 達上野 一樹池野 正弘田中 真伸内田 智久
著者情報
ジャーナル フリー

2013 年 23 巻 3 号 p. 218-222

詳細
抄録

We have developed a prototype time-resolved neutron imaging detector employing a micro-pattern gaseous detector known as the micro-pixel chamber (μPIC) coupled with a field-programmable-gate-array-based data acquisition system. Our detector system combines 100μm-level spatial and sub-μs time resolutions with a low gamma sensitivity of less than 10-12 and high data rates, making it well suited for applications in neutron radiography at high-intensity, pulsed neutron sources. In the present paper, we introduce the detector system and present several test measurements performed at NOBORU (BL10), J-PARC to demonstrate the capabilities of our prototype. We also discuss future improvements to the spatial resolution and rate performance.

著者関連情報
© 2013 日本中性子科学会
前の記事 次の記事
feedback
Top