J-PARC ANNRIビームラインにおける偏極熱外中性子利用
遠藤 駿典奥平 琢也
著者情報
ジャーナル フリー

2023 年 33 巻 2 号 p. 68-72

詳細
抄録

Accurate Neutron-Nucleus Reaction measurement Instrument (ANNRI) is installed at the one of the beamlines in MLF in J-PARC, and nuclear reaction measurements have been conducted using the three types of γ-ray detectors, Ge, NaI, and LaBr. Recently, a 3He spin filter to polarize neutrons has become available in ANNRI, and several experiments have been performed using the produced polarized neutrons. Significant differences in the measured cross section and γ-ray emission depending on the direction of the neutron polarization were observed in the 117Sn(n,γ)118Sn reaction. A γ-ray polarimeter was also developed, and the circular polarization of emitted γ-rays was successfully measured at ANNRI for the 32S(n, γ)33S reaction using the polarized neutrons.