放射線
Online ISSN : 2758-9064
特集 「中性子の線量評価に関わる研究開発」
中性子捕捉療法のための熱中性子リアルタイムモニタの開発
石川 正純高橋 浩之小野 公二櫻井 良憲熊田 博明町島 祐一
著者情報
ジャーナル フリー

2005 年 31 巻 4 号 p. 279-285

詳細
抄録

  A wide range thermal neutron detector was developed based on the Scintillator with Optical Fiber (SOF) detector which has been previously used for thermal neutron monitoring during boron neutron capture therapy irradiation. With this new detector system we intended to address the issues of real-time thermal neutron flux measurement and the simultaneous measurement of a wide range of thermal neutron flux in a BNCT irradiation field which were difficult to implement with the gold wire activation method. Clinical trials using paired SOF to measure thermal neutron flux during therapy confirmed that paired SOF detectors were effective as a real-time thermal neutron flux monitor.

著者関連情報
© 2005 本論文著者