放射線
Online ISSN : 2758-9064
特集 「中性子の線量評価に関わる研究開発」
中性子捕捉療法のための線量評価システム、JCDSの開発
熊田 博明山本 和喜松村 明中川 義信
著者情報
ジャーナル フリー

2005 年 31 巻 4 号 p. 287-297

詳細
抄録

  Clinical trials for boron neutron capture therapy (BNCT) are performed using research reactors in world-wide. BNCT is a radiation therapy for refractory cancer as malignant brain tumor. To carry out the BNCT procedure based on proper treatment planning and its precise implementation, a treatment planning system, JCDS has been developed in Japan Atomic Energy Research Institute (JAERI; now restructured to Japan Atomic Energy Agency, JAEA).

  To estimate the dose components given to a patient by neutron irradiation in BNCT, JCDS performs the dosimetry with numerical simulations. A detailed tree-dimensional patient's head model is created by using CT and MRI images, and then input data for neutron and photon transport calculation are generated based on the patient's model. Distributions of the several dose components and gamma-ray dose in the head are computed by Monte Carlo calculations. A proper irradiation condition based on the calculation results is decided, and then a clinical trial of BNCT can be performed.

This report describes requirements for dosimetry in BNCT and the dosimetry method of JCDS.

著者関連情報
© 2005 本論文著者
前の記事 次の記事
feedback