皮膚ガン診断用サーミスタ特性校正に用いる真空断熱式温度校正装置の評価
藤田 法明野中 崇細川 靖古川 琢磨郭 福会井関 祐也横田 実世岡部 孝裕圓山 重直
著者情報
ジャーナル 認証あり

2024 年 144 巻 4 号 p. 126-131

詳細
抄録

Recent methods using thermistors as an early detection of skin cancer require the higher resolution of 5 mK. Previously, the authors developed a vacuum insulated temperature device that calibrates the resistance temperature characteristics of thermistors, so that it improves the temperature resolution and accuracy of skin cancer diagnosis. In this study the operational methods of the calibration system were compared for improving further accuracy using the temperature calibration device and the oil tank calibration device. The results of the comparison indicated that the temperature calibration device is probably as accurate as the oil tank calibration device.