International Heart Journal
Online ISSN : 1349-3299
Print ISSN : 1349-2365
ISSN-L : 1349-2365
Clinical Studies
MRI Mode Programming for Safe Magnetic Resonance Imaging in Patients With a Magnetic Resonance Conditional Cardiac Device
Toshiko NakaiSayaka KurokawaYukitoshi IkeyaKazuki IsoKeiko TakahashiNaoko SasakiSonoko AshinoKimie OkuboYasuo OkumuraSatoshi KunimotoIchiro WatanabeAtsushi Hirayama
著者情報
ジャーナル フリー

2016 年 57 巻 2 号 p. 173-176

詳細
抄録
Although diagnostically indispensable, magnetic resonance imaging (MRI) has been, until recently, contraindicated in patients with an implantable cardiac device. MR conditional cardiac devices are now widely used, but the mode programming needed for safe MRI has yet to be established. We reviewed the details of 41 MRI examinations of patients with a MR conditional device. There were no associated adverse events. However, in 3 cases, paced beats competed with the patient’s own beats during the MRI examination. We describe 2 of the 3 specific cases because they illustrate these potentially risky situations: a case in which the intrinsic heart rate increased and another in which atrial fibrillation occurred. Safe MRI in patients with an MR conditional device necessitates detailed MRI mode programming. The MRI pacing mode should be carefully and individually selected.
著者関連情報
© 2016 by the International Heart Journal Association
前の記事 次の記事
feedback
Top