International Journal of Automation Technology
Online ISSN : 1883-8022
Print ISSN : 1881-7629
ISSN-L : 1881-7629
Special Issue on Laser Metrology for Precision Measurement and Inspection in Industry
Development of Dimensional X-Ray Computed Tomography
Hiroyuki FujimotoMakoto AbeSonko OsawaOsamu SatoToshiyuki Takatsuji
著者情報
ジャーナル オープンアクセス

2015 年 9 巻 5 号 p. 567-571

詳細
抄録

Recently, a strong need has arisen for a dimensional X-ray computed tomography system that is capable of dimensional measurements. This is because the speedy realization of dimensional measurements for outward forms and inward forms on dense spatial points remarkably simplifies and accelerates production loop. However, although the image obtained via XCT describes the structure clearly and in great detail, dimensional metrology by means of XCT is not simple. The National Metrology Institute of Japan has been carrying out performance tests using gauges that include the gauges proposed in ISO10360. In this work, the magnification variation correction is carefully presented, and a maximum deviation of less than 5 μm is shown to be possible by means of the measurement of the forest phantom of 27 ruby spheres, the locations of which are calibrated by the coordinate measuring machine.

著者関連情報

この記事は最新の被引用情報を取得できません。

© 2015 Fuji Technology Press Ltd.

This article is licensed under a Creative Commons [Attribution-NoDerivatives 4.0 International] license (https://creativecommons.org/licenses/by-nd/4.0/).
The journal is fully Open Access under Creative Commons licenses and all articles are free to access at IJAT Official Site.
https://www.fujipress.jp/ijat/au-about/
前の記事 次の記事
feedback
Top