2020 Volume 69 Issue 4 Pages 615-622
The serum-soluble interleukin-2 receptor has been used as a diagnostic aid for malignant lymphoma and as an indicator of treatment follow-up. In this study, we compared the performance of reagents by chemiluminescent enzyme immunoassay (CLEIA; Lumipulse Presto IL-2R, manufactured by Fujirebio Co., Ltd.) and latex turbidimetric immunoassay (Nanopia IL-2R, manufactured by Sekisui Medical Co., Ltd.). Furthermore, we compared the correlation of CLEIA using these reagents with the outsourced assay (Determiner CL IL-2R, Hitachi Chemical Diagnostics Systems Co., Ltd.) and evaluated the suitability of these reagents in in-hospital measurement. The performance of both reagents was satisfactory and sufficient for daily measurements. However, since Nanopia IL-2R showed nonspecific reactions in some samples, it is desirable to use it considering the occurrence of nonspecific reactions. It is possible to immediately select the appropriate treatment method and expect a clinical diagnostic contribution by reporting the results on the same day.