1978 年 34 巻 10 号 p. 563-568
This paper presents a new modulation method for an ultrasonic Doppler flow meter with high range resolution. In this method, transmitted and received waves are modulated and demodulated respectively depending upon the specified new sequences. The cross-correlation values are zero except for peak value. Is sought a pairof new sequences with such cross-correlation property which provides high sensitivity for detection. The advantages of the proposed method are that it employes continuous waves and enables complete elimination of unfavorable reflected waves. Therefore, the present method is superior to a pulsed Doppler method and an M-sequence modulation method owing to higher detection sensitivity.