JOURNAL OF CHEMICAL ENGINEERING OF JAPAN
Online ISSN : 1881-1299
Print ISSN : 0021-9592
A MATHEMATICAL MODEL FOR TRANSFER OF O2 AND CO2 IN A MEMBRANE ARTIFICIAL LUNG
OSCAR SOSAJOSÉ C. MERGHUK
Author information
JOURNAL FREE ACCESS

1974 Volume 7 Issue 1 Pages 25-30

Details
Abstract
An analysis of oxygen absorption and carbon dioxide desorption in a film of blood through a gas-permeable membrane is presented. The mathematical analysis treats blood as a homogeneous, non-Newtonian fluid, in which the system pO2, pCO2 hemoglobin, pH is always in equilibrium due to the reaction rate in the erythrocites.
Influence of flow rate, hematocrit and gas composition on blood saturation are studied, and the possibility of calculating an optimal membrane length is found.
Content from these authors
© The Society of Chemical Engineers, Japan
Previous article Next article
feedback