希釈冷凍機の現状
児玉 隆夫信貴 豊一郎
著者情報
ジャーナル フリー

1971 年 6 巻 4 号 p. 137-145

詳細
抄録

Following the brief description of the basic principles of dilution cooling briefly, recent developments of dilution refrigeration are reviewed. In order to realize the idea of dilution refrigeration as perfectly as possible, following factors are important; to make the efficient heat exchangers which have low flow-impedance and small thermal boundary resistance, to reduce the amount of He4 in the circulating He3 and to eliminate heat input into the mixing chamber, especially vibration of the cryostat originating in mechanical pumps.
Three types of heat exchangers-continuous or tube-in-tube heat exchanger, sintered copper heat exchanger and copper foil heat exchanger are described.
Almost pure He3 can be circulated by use of the novel design of still which suppresses the He4 film-flow.
The lowest temperature which has been achieved so far by the dilution refrigerator is 5.5mK in the continuous operation and 4.5mK in the singlecycle operation. This value is very close to the theoretical limiting temperature of the dilution refrigerator.
Experiments which combine the continuous dilution refrigeration with the adiabatic demagnetization and the adiabatic compression of He3 are also presented.