抄録
In this paper, thé author has presented the Temperature Programed Retention Indices (PTRI) of 1, 340 environmental chemicals on methyl silicone capillary column. PTRI is defined as follows:
PTRI =100×Z +100× (TX-TZ) / (Tz+1-TZ)
Z: Carbon number of n-alkane eluted just before the target compound.
TX, TZ and TZ+1: Retention times of the target compound (x), n-alkane of z and z+1 carbon munber, respectively.
The author expects these data are applied to identifications of environmental chemicals using GC/MS.