Development of Hydrocarbon Flow Calibration Facility as a National Standard
Takashi SHIMADARyouji DOIHARAYoshiya TERAOMasaki TAKAMOTO
著者情報
ジャーナル フリー

2007 年 2 巻 1 号 p. 23-34

詳細
抄録

A new primary standard for hydrocarbon flow measurements has been constructed at National Metrology Institute of Japan (NMIJ). The facility was designed for the calibration of hydrocarbon flowmeters in the flow rate range between 3 and 300 m3/h. The expanded uncertainty is estimated to be 0.03 % for volumetric flow rate and 0.02 % for mass flow rate (coverage factor: k = 2). The primary standard is based on a static and gravimetric method with a flying start and finish. The facility consists of two test rigs using kerosene and light oil as working fluids. The test lines for the flowmeters are 50, 100 and 150 mm in diameter and three servo positive displacement meters are used as working standards. To verify the calibration performance, a Coriolis flowmeter, a turbine meter and a positive displacement flowmeter have been calibrated at both test rigs. Furthermore, an international comparison with SP, Swedish National Testing Research Institute, was carried out. A screw-type positive displacement flowmeter was selected as the transfer standard and was calibrated at NMIJ and SP. The result shows that the two national standards at the two institutes agree within the quoted expanded uncertainties.