電離箱型およびGM計数管型サーベイメータの校正 (I)
作業環境のβ線線量率測定のための試験
備後 一義梶本 与一須賀 新一
著者情報
ジャーナル フリー

1978 年 13 巻 4 号 p. 279-285

詳細
抄録
Three types of ionization chamber survey meters and a type of GM counter survey meter were calibrated for measuring the β-ray absorbed dose rate in a working area. To estimate the β-ray absorbed dose rate, a survey meter was used without and with a filter. A reading of survey meter's indicator measured with the filter was subtracted from a reading measured without the filter, and then the absorbed dose rate was obtained by multiplying this remainder by a conversion coefficient.
The conversion coefficients were roughly constant with distance more than 8cm (ionization chamber survey meters) and with distance more than 5cm (GM counter survey meter). The conversion coefficient was dependent on β-ray energies. In order to measure the absorbed dose rate of tissue whose epidermal thickness is 40mg/cm2, the constant value, 4(mrad/h)/(mR/h), was chosen independently of β-ray energies as the conversion coefficient of three types of ionization chamber survey meters. The conversion coefficient of the GM counter survey meter was more energy dependent than that of every type of ionization chamber survey meter.