日本歯科麻酔学会雑誌
Online ISSN : 2433-4480
臨床報告
静脈留置針に製造過程において生じた亀裂を認めた2症例
坂野上 和奏柳町 晴香前澤 五月伊藤 佳奈大野 真由子佐藤 健一
著者情報
ジャーナル フリー

2025 年 53 巻 2 号 p. 60-63

詳細
抄録

  Intravenous indwelling needles are widely used in intravenous sedation and perioperative management for general anesthesia. We report two cases in which cracks were observed near the hub of indwelling needles. The first patient was a 47-year-old woman with dental phobia who was scheduled to receive dental treatment under intravenous sedation. Peripheral intravenous access was secured using the BD InsyteTM AutoguardTM BC Pro 24G, manufactured by Becton, Dickinson and Company. Immediately after initiating the infusion, fluid leakage was observed near the connection between the intravenous needle and the IV circuit. The intravenous needle was removed, and another intravenous needle was used to resecure the venous access. The second patient was a 39-year-old woman who was scheduled to undergo an intramaxillary foreign body removal after sagittal splitting of the mandibular ramus under intravenous sedation. An intravenous line was established using the above-mentioned product from the same company and an infusion was initiated ; immediately afterwards, however, the infusion began to leak from the indwelling needle. In both cases, the intravenous access was resecured, and intravenous sedation was performed without any subsequent problems. When the removed indwelling needles were examined, cracks were found near the hub of the indwelling needles in both cases. Safety inspections are necessary before using medical devices, but unexpected problems can occur. In such cases, both early problem recognition and appropriate responses are necessary.

著者関連情報
© 2025 一般社団法人日本歯科麻酔学会
前の記事 次の記事
feedback
Top