2022 Volume 29 Issue 3 Pages 164-178
As basic research for contributing to the determination of conformity with the Positive List (PL) system, chemical substances contained in plastic utensils, containers and packaging were comprehensively analyzed using GC-MS. Total of 131 standard substances of additives listed in the Japanese PL were analyzed to obtain information on mass spectra and retention index (RI). The ratio of calculated RI to library-registered RI showed a good agreement, being 1.0 for most substances. By comparing RI in addition to mass spectra, unknown peaks detected in plastic products could be identified more reliably. Chemical substances contained in plastic products made of polypropylene, polyethylene, polystyrene, and polyethylene terephthalate, which are widely used generalpurpose plastics, were analyzed. As a result, 61 of the 104 substances detected in plastic products were identified and/or estimated, and about half of them were the ones listed in PL. However, the other half were non-listed substances, and most of them were considered to be non-intentionally added substances (NIAS), which are degradation products, impurities, by-products, etc. of plastic and additives.