医学物理
Online ISSN : 2186-9634
Print ISSN : 1345-5354
ISSN-L : 1345-5354
研究論文
残留飛程分布測定に基づく重イオンCTの高速撮影法の検討
茂垣 達也阿部 慎司佐藤 斉村石 浩原 秀剛原 敏三宅 晶子日向 猛金井 達明
著者情報
ジャーナル フリー

2012 年 32 巻 2 号 p. 58-66

詳細
抄録
We proposed a new technique for the fast acquisition heavy ion CT (HICT) system based on the measurement of residual range distribution using an intensifying screen and charge coupled device cam-era. The previously used fast acquisition HICT system had poor electron density resolution. In the new technique, the range shifter thickness is varied over the required dynamic range in the spill of the heavy ion beam at each projection angle and the residual range distribution is determined by a series of acquisi-tion data. We examined the image quality using the contrast noise ratio and the noise power spectrum, and estimated the electron density resolution, using a low-contrast phantom for measurement of electron density resolution. The image quality of the new technique was superior to that of the previous fast ac-quisition HICT system. Furthermore, the relative electron density resolution was 0.011, which represent-ed an improvement of about 12-fold. Therefore we showed that the new technique was potentially useful in clinical use of HICT, including treatment and quality assurance of heavy ion therapy.
著者関連情報
© 2012 日本医学物理学会
前の記事 次の記事
feedback