グラファイトカロリメータによる水吸収線量率の一次標準
森下 雄一郎
著者情報
ジャーナル フリー

2013 年 33 巻 4 号 p. 179-185

詳細
抄録
The calibration service in terms of absorbed dose to water started from 2011 after establishment of the national primary standard using a graphite calorimeter at the national metrology institute of Japan (NMIJ) and JCSS accreditation of the association for nuclear technology in medicine (ANTM). Accordingly, a new dosimetry protocol was introduced as JSMP12, in which details of the national standard were also described. This report presents a short review of the standard, a key comparison result, and a comparison result of calibration coefficients by JSMP01 and JSMP12.
著者関連情報
© 2013 日本医学物理学会
前の記事 次の記事