医学物理
Online ISSN : 2186-9634
Print ISSN : 1345-5354
ISSN-L : 1345-5354
解説
線質変換係数
河内 徹片寄 哲朗
著者情報
ジャーナル フリー

2013 年 33 巻 4 号 p. 186-191

詳細
抄録
This report describes the update of the beam quality conversion factor KQ,Q0 of the standard dosimetry protocol in Japan. The KQ,Q0 corrects for the difference between the response of an ionization chamber in the reference beam quality Q0 used for calibrating the chamber and in the actual user beam quality Q. All changes of KQ,Q0 were caused by the perturbation correction factors which were recalculated by Monte Carlo simulation. With a calculation process, unsolved problems in this update are also discussed here.
著者関連情報
© 2013 日本医学物理学会
前の記事 次の記事
feedback
Top