医学物理
Online ISSN : 2186-9634
Print ISSN : 1345-5354
ISSN-L : 1345-5354
解説
マンモグラフィの線量の国家標準の進展と展望
田中 隆宏
著者情報
ジャーナル フリー

2021 年 41 巻 3 号 p. 87-91

詳細
抄録

In Japan, mammography was introduced in 2000 for the early detection of breast cancer. In quality control of mammography, dosimetry is one of the most important items. For accurate dosimetry, calibration of dosimeters is necessary because radiation quality (target-filter combination) of mammography x-ray is different from that of general radiography. Therefore, development of dosimetry standard based on radiation quality of mammography x-ray was required. AIST/NMIJ developed an air-kerma standard for mammography x-rays and started its dissemination in 2009. Since then, the air-kerma standard has been extended to various radiation qualities that have come to be used in digital mammography. In this paper, an overview of the air-kerma standard for mammography x-ray together with a future plan is briefly presented.

著者関連情報
© 2021 日本医学物理学会
前の記事 次の記事
feedback
Top