フラットパネルディテクタシステムの物理的画像特性
吉田 彰中村 悟西原 貞光小濱 千幸高畑 明藤川 光一
著者情報
ジャーナル フリー

2002 年 22 巻 4 号 p. 246-254

詳細
抄録
We report the physical imaging properties of a flat panel detector(FPD) designed for radiographic imaging applications(Revolution XQ/i digital chest imaging system, G. E. Medical Systems). The imaging properties of the detector were evaluated through measurements of the characteristic curve, modulation transfer function(MTF), and Wiener spectrum.
The digital characteristic curves of the FPD system were measured for the two tube voltages(80kV and 120kV). They showed that the correlation between the pixel values of FPD and the incident exposure to the FPD was a linear correlation. The dynamic range of characteristic curves had a range from about 0.003 to 2 μ C/kg at the exposure. The presampling MTFs for different tube voltage were almost the same, when we measured the MTFs at 80kV and 120kV. And also, there was no significant difference between the MTFs measured with the slit in the direction parallel to the horizontal direction and with the slit in the perpendicular direction. The relative resolution of the FPD system was a 46% higher than that of ST V imaging plate of FCR(Fuji Computed Radiography) system at 2 cycles/mm. The digital Wiener spectrum of the FPD system was about 1/10 lower than that of the CR system. Thus the FPD system can provide superior imaging performance due to both high resolution and low noise.