FPD搭載型血管造影装置の臨床経験
奥迫 謙治庄垣 雅史横山 貢治小川 隆由田中 克尚市田 隆雄岸本 健治畑川 政勝奥山 和夫工藤 弘明田中 佐織中村 健治池田 重之鈴木 克巳
著者情報
ジャーナル フリー

2002 年 22 巻 4 号 p. 255-263

詳細
抄録
We had the chance of clinical studies by new proto-type Digital Subtraction Angiography (DSA) System with the Flat Panel Detector (FPD).
That system has two types of detector, FPD and I. I. -CCD camera, and we can compare with image quality in clinical.
We measured and discussed about the physical characteristic data of both detectors. Up to physical characteristic evaluation, FPD has good MTF and SNR performance, comparing with I.I.-CCD camera. In clinical evaluation, DSA image of FPD was superior to that of I. I. -CCD camera by visual comparison of medical doctors. Finally we discussed about the feasibility of replacement of detectors. We expect that DSA system will accept FPD in near future.