放射線治療施設での治療用加速器の出力線量に関する訪問調査(埼玉県および栃木県内での調査結果)
榎戸 義浩岡本 裕之清宮 幸雄古谷 智久藤森 健吾草野 陽介遠山 尚紀伍賀 友紀大山 正哉峯村 俊行中村 譲新保 宗史田伏 勝義石倉 聡遠藤 真広池田 恢
著者情報
ジャーナル フリー

2007 年 27 巻 2 号 p. 71-77

詳細
抄録
Field survey on output for X-ray therapeutic accelerators took place three times in Saitama Prefecture. The result of the field survey in 1997 showed the different rate from the designated dose at peak depth of 35 beams in 18 institutions. As different rate within ±-5% stood 91.4% in all beams, so different rate within ±3% stood 85.7% in the same beams. The average different rate from the designated dose at peak depth was +1706%. The standard deviation of the same condition was 3.72.
The result of the field survey in 2005 showed the different rate from the designated dose at correction depth of 36 beams in 18 institutions. As different rate within ± 5% stood 100% in all beams, so different rate within ±3% stood 91.6% in the same beams. The average different rate from the designated dose at correction depth was +0.80%. The standard deviation of the same condition was 1.46.
We und e rstood that the different rate from the designated dose at radiotherapeutic institutions decreased and even the value of the standard deviation was decreasing, by receiving 3 times of field surveys that was held in Saitama Pref. Also we understood that the beam numbers of different rate within ±5% and the beam numbers of different rate within ±3% were going up. We recognized that the good result of accurate dose is obtained more, by doing a continual field survey. The field survey was carried out in 2006 in Tochigi Prefecture and was the insufficient result in 10% of institutions.
著者関連情報
© 一般社団法人 日本医学物理学会
前の記事
feedback
Top