医療薬学
Online ISSN : 1882-1499
Print ISSN : 1346-342X
ISSN-L : 1346-342X
一般論文
蛍光ラベル化剤を用いたHPLC による小児血漿中トピラマート濃度測定法の開発
高橋 良平今井 浩一吉田 栄充田中 学浜野 晋一郎新井 康俊
著者情報
ジャーナル フリー

2010 年 36 巻 7 号 p. 486-492

詳細
抄録
In this study,we developed a reversed-phase high-performance liquid chromatographic method using a fluorescence derivatizing agent for the determination of topiramate (TPM) in infant plasma.The method used 100μL plasma samples,and the HPLC separation was carried out on a Discovery HS-C 18 column (3μm,4.6 mm×150 mm) with phosphate buffer (pH 2.5 ; 25 mM)-methanol-acetonitrile (50 : 40 : 10 (v/v/v)) as the mobile phase at a 1.0 mL/min flow rate.TPM was detected using a fluorescence detector (λex/em : 470/537 nm).The quantification limit was 0.1μg/mL,and the intra- and interassay coefficients of variation were 0.8-7.1% and 1.3-9.3%,respectively.A specificity and selectivity study showed that there were no interfering peaks.Furthermore,correlation analysis of TPM concentrations measured by HPLC and LC/MS produced a correlation coefficient (γ) of 0.9825.
著者関連情報
© 2010 日本医療薬学会
前の記事 次の記事
feedback