輸液セットの装着手技が流量制御式輸液ポンプの投与速度の精度に及ぼす影響
梅村 雅之大野 裕之重野 克郎畠山 和人矢野 亨治森 智子脇屋 義文
著者情報
ジャーナル フリー

2013 年 39 巻 8 号 p. 459-464

詳細
抄録
Accurate continuous intravenous injection via peristaltic finger infusion pump has been utilized at outpatient clinics, such as cancer treatment centers. However, the flow rate accuracy of the peristaltic finger infusion pump might be affected by various factors, such as the properties of drugs containing surfactants, choice of infusion set, and the technique for using the peristaltic finger infusion pump by medical staff. In our experience with outpatient chemotherapy, medication administration has occasionally been incomplete at the calculated input time when using peristaltic finger infusion pump. For this reason, we have investigated the cause of delay in administration time with peristaltic finger infusion pump under various conditions. Saline containing ethoxylated hydrogenated castor oil (cremophor® EL), polyoxyethylene sorbitan monooleate (Tween 80), and ethanol did not adversely affect administration time accuracy. Conversely, when the infusion set tubing was stretched approximately 1-3 cm and was attached to the peristaltic finger infusion pump, the time required for complete administration of the solution was prolonged compared to the input time (P < 0.0001-0.05, Tukey-Kramer multiple comparison). Therefore, we suggest that the technique for using the peristaltic finger infusion pump by the medical staff adversely prolonged the time required for completion of administration. In our opinion, pharmacists should provide information concerning not only drugs, but also medical devices to physicians and nurses.
著者関連情報
© 2013 日本医療薬学会
次の記事
feedback
Top