抄録
The fluorescence polarization immunoassay (FPIA) method for the determination of serum aprindine (AP) concentrations was evaluated by comparison with HPLC method.The withinrun precision of FPIA method was 4.6-6.8% as the coefficient of variation (C. V.).The C.V. value of between-run precision was 4.1-8.7%.The values of serum AP concentrations determined by FPIA method correlated well with those determined by HPLC method.The correlation coefficient was 0.987.FPIA method, reliable, simple and rapid, seems to be a useful method for routine clinical use.