Protein Effect on the Antimony Microelectrode in Application to Biological Fluid
佐竹 典子松村 裕藤本 守
著者情報
ジャーナル フリー

1980 年 30 巻 5 号 p. 689-700

詳細
抄録
In order to examine the effect of protein on the Sb-microelectrode used for the pH measurement of biological fluids, bovine serum albumin (BSA) and plasma protein were used for test samples. It was found that, BSA, probably by binding to components in the electrode system, produced a marked deviation of both the electromotive force and slope constant of the microelectrode, resulting in an alkaline shift of pH of 0.1-0.7. But the shift of the pH reading could be empirically predicted within an error of±0.03 from the cubic function of protein concentration.Use of Tris standard solution as the calibration buffer, together with corrections for the effects of temperature, ionic strength, PO2, etc. on the Sb-microelectrode, would reduce the alkaline shift to approximately less than 0.08 and 0.04 pH units in the pH measurement of blood plasma and whole blood, respectively.