2018 年 17 巻 1 号 p. 2-8
It is important to measure scattered radiation in diagnostic radiology in terms of evaluation of occupational exposure and leakage radiation dose. Recently, Unfors RaySafe (Sweden) released a solid-state survey sensor aimed at expanding RaySafe X2 system. However, there are no detailed reports regarding the performance, which includes energy dependence and angular dependence, of X2 survey meter. The purpose of this study was to evaluate the performance of X2 survey meter. Experiments were conducted to compare the performance of X2 survey meter with that of an ionization chamber survey meter. The results showed that the fundamental performance of X2 survey meter were equal to or superior to that of an ionization chamber survey meter. In addition, X2 survey meter is easy to handle. The study concludes that X2 survey meter has adequate performance for measurement of scattered radiation in diagnostic radiology.