プラスチック短下肢装具歩行立脚相における荷重ベクトルの状態と装具底背屈角度との関係
北野 将利北山 一郎森岡 大輔山中 隆池原 用祐小山 秀世森本 高志園部 秀樹宮崎 展行
著者情報
ジャーナル フリー

2016 年 16 巻 2 号 p. 147-153

詳細
抄録
   Plastic ankle-foot orthoses (PAFOs) are often used as assistive devices for patients with hemiplegia. In this study, force exerted on PAFOs during walking was measured using a sensor installed at the top of the calf region on the PAFO, where the structure of the PAFO exerts minimal influence on measurements, and the data were analyzed. First, PAFOs with incorporated sensors were individually prepared for 3 healthy individuals and for 3 patients with hemiplegia. Mechanical data were then obtained while individuals wore the PAFOs during level-ground walking. We found that at the early stance phase, force vectors apply at a slightly lower region of the prosthetic rear wall, the level of which is the same as the upper belt, and at the late stance phase, they usually apply at the upper belt region. We also found that the waveform of posterior force applied by the PAFOs coincide with that of the bending angle of the PAFOs. From these findings, we found that the approximate posterior direction force applied by the PAFOs, which usually must be measured using costly sensors and modified PAFOs, can be easily estimated by measuring and calculating from PAFO bending angle.