抄録
The wideband measurement system of permeability for magnetic thin slab was proposed for up to 10 GHz, using the all shielded shorted microstrip lines. In order to analyze this configuration, the reference plane was set on the other end of specimen where the LC parallel resonant circuit was assumed. The permeability was derived from the measurement of S11 by VNA (Vector Network Analyzer) comparing between with strong static magnetic field parallel to rf magnetic field and without it. The shorted microstrip line was calibrated by the standard sample of rectangular shape NSS (Noise Suppression Sheet) comparing with the results of toroidal core measured by the shorted coaxial line.