Journal of Light & Visual Environment
Online ISSN : 1349-8398
Print ISSN : 0387-8805
ISSN-L : 0387-8805
Research Note
Domestic Comparison of Relative Spectral Responsivity Measurements for Illuminance Meters
Kazuaki OHKUBOMotomi HORIUCHIYasuo NAKAGAWAHitoshi TOZAWAKenji KOBAYASHIIzumi HORlENaomich’ CHIDA
著者情報
ジャーナル フリー

2000 年 24 巻 1 号 p. 1_66-1_72

詳細
抄録
A comparison of relative spectral responsivity measurements of illuminance meters was carried out among five companies — TOPCON Corp., HIOKI E.E. Corp., Minolta Co., Ltd., Yokogawa M&C Corp., and Matsushita Electric Industrial Co., Ltd. The spectral responsivitys of the illuminance meters were measured by each company's spectral responsivity measurement system calibrated to each company's standard. As a result, the measured values of fs (fs is defined as the relative spectral responsivity error in Japanese Industrial Standard JIS-C-1609) differed within 1.7. The fs is identical with the f1' recommended by the CIE.
著者関連情報
© 2000 The Illuminating Engineering Institute of Japan
前の記事
feedback
Top