Semiautomatic Treatment Planning for the Field-in-field Technique in Whole Brain Irradiation
HIROYUKI WATANABE SATORU SUGIMOTOTORU KAWABATAHIRONORI NAGATACHIE KUROKAWAKEISUKE USUITATSUYA INOUEJUN TAKATSUKYOICHI KATOKEISUKE SASAI
Author information
JOURNAL OPEN ACCESS Advance online publication

Article ID: JMJ22-0003-OA

Details
Abstract

Objectives In radiation therapy, the field-in-field (FIF) technique is used to prevent the administration of unnecessarily high doses to reduce toxicity. Recently, the FIF technique has been used for whole brain irradiation (WBI). Using the FIF technique, the volume that receives a higher than prescribed dose (hotspot) can be largely reduced; however, the treatment planning requires time. Therefore, to reduce the burden on the treatment planners, we propose a semiautomatic treatment planning method for the FIF technique.

Methods In the semiautomatic FIF technique, hotspot regions in a treatment plan without the FIF technique are identified three-dimensionally, and beams with blocks that cover the hotspot regions using a multileaf collimator (sub-beams) are automatically created. The sub-beams are added to the original plan, and weights are assigned based on the maximum dose of the original plan to decrease the doses in the hotspot regions. This method was applied to 22 patients previously treated with WBI, wherein treatment plans were originally created without the FIF technique.

Results In the semiautomatic FIF plans, the hotspots almost disappeared. The dose to 95% of the volume and the volume receiving at least 95% of the prescribed dose in the planning target volume decreased by only 0.3% ± 0.2% and 0.0% ± 0.1%, respectively, on average compared with those in the original plan. The average semiautomatic FIF processing time was 28 ± 4 s.

Conclusions The proposed method reduced the hotspot regions with a slight change in the target coverage.

Content from these authors
© 2022 The Juntendo Medical Society. This is an open access article distributed under the terms of Creative Commons Attribution License (CC BY), which permits unrestricted use, distribution, and reproduction in any medium, provided the original source is properly credited.

This article is licensed under a Creative Commons [Attribution 4.0 International] license.
https://creativecommons.org/licenses/by/4.0/