Journal of Neuroendovascular Therapy
Online ISSN : 2186-2494
Print ISSN : 1882-4072
ISSN-L : 1882-4072

この記事には本公開記事があります。本公開記事を参照してください。
引用する場合も本公開記事を引用してください。

Atypical Incomplete Detachment Following PulseRider Deployment
Shunsaku GotoTakashi IzumiMasahiro NishihoriYoshio ArakiKinya YokoyamaKenji UdaRyuta Saito
著者情報
キーワード: PulseRider, complication, case report, rare
ジャーナル オープンアクセス 早期公開

論文ID: cr.2021-0095

この記事には本公開記事があります。
詳細
抄録

Objective: Owing to the limited time since the introduction of the PulseRider (PR), inconsequential or rare complications that clinicians should be aware of remain unreported yet. Here, we report a rare complication of incomplete detachment.

Case Presentation: A 50-year-old male underwent PR-assisted coil embolization for a basilar tip aneurysm. Coiling was completed, and the detachment procedure was performed using a detachment machine; the success signal was observed. The delivery microcatheter was subsequently advanced back up to the proximal markers, and no reapproximation of the proximal markers, which indicates successful detachment, was observed. However, only one of the proximal markers returned to the microcatheter, and incomplete detachment of only one leg was detected. Ultimately, electrical detachment was not possible, and physical separation by tension was achieved.

Conclusion: Our case report presents a rare case of a detachment problem in the PR. The PR could not be detached, although the signal revealed successful detachment. Therefore, careful withdrawal of the delivery wire by checking not only the proximal markers but also the behavior of the entire PR and coil complex is important.