Journal of Nuclear Science and Technology
Online ISSN : 1881-1248
Print ISSN : 0022-3131
Development of Position Sensitive Helium-3 Proportional Counter for Thermal Neutrons with Novel Signal Readout
Akira URITANIChizuo MORITamaki WATANABEAkira MIYAHARA
著者情報
ジャーナル フリー

1990 年 27 巻 8 号 p. 712-717

詳細
抄録
A position sensitive 3He proportional counter for thermal neutrons with a novel signal readout method which has two pre-amplifiers at only one end of the counter body has been developed. It has an advantage for usage in a special environment, such as in a subcritical facility or in water, or for insertion into a narrow space. The characteristics of the counter were evaluated with thermal neutrons. The position resolution was 7 mm with an effective length of 1, 200 mm. The integral non-linearity of position signal was less than 0.5%.
著者関連情報

この記事は最新の被引用情報を取得できません。

© the Atomic Energy Society of Japan
前の記事 次の記事
feedback